ARL Biopharma, Inc. dba Analytical Research Laboratories Form 483, November 2012
FDA issued this Form 483 to ARL Biopharma, Inc. dba Analytical Research Laboratories in Oklahoma City, Oklahoma (FEI 3003644883) at the end of a drug inspection that ended November 8, 2012. It lists 4 observations. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Nov 8, 2012
- Drugs · Drug Quality Assurance
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Jan 17, 2024
- 4,087 days after the inspection
The 4 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Incomplete USP <71> sterility practices
1. Your firm states on the Microbiology Report that is issued to a client after sterility and/or fungal testing that the Test Method employed was USP <71>. However, your firm is not fully following all parts of USP <71> when performing sterility and/or fungal testing of human drug products. For example, a. USP <71> requires a Method Suitability Test be performed for all new products tested. Your firm does not have documentation to show that Method Suitability Testing has been performed for all drug products submitted for sterility testing by Ameridose, LLC and New England Compounding Center (NECC), both located in Massachusetts. For those drug products submitted by NECC and Ameridose, LLC, you have some documentation of bacteriostasis/fungistasis testing performed in 2006 & 2008 on a limited number of drug products, however there is no source documentation showing how the tests were performed, lot numbers of organisms or media used, and who performed the testing. b. USP <71> specifies the number of articles to be tested. While you provide reference to USP <71> for sample sizes, you do not ensure that your clients are submitting the required number of articles for testing. Most clients usually submit only [illegible] for sterility testing, including NECC and Ameridose.
- 2
Unvalidated potency (assay) methods
2. Your firm has no documentation to show that all analytical methods used to test for potency (assay) have been validated for all drug products including drug products submitted for testing by NECC and Ameridose LLC. These include drug products such as Methylprednisolone Acetate, Heparin, Vasopressin, Triamcinolone Acetonide, and products containing Bupivacaine and Epinephrine. Analytical methods that are not validated and/ or not found in the USP that are used for potency testing of human drug products are not written, reviewed and approved.
- 3
Incomplete USP <85> endotoxin procedures
3. Your firm states on the Microbiology Report that is issued to a client after endotoxin testing that the Test Method employed was USP <85>. Your firm is not fully following all parts of USP <85> when performing endotoxin testing of human drug products. Specifically, the Maximum Valid Dilution (MVD) is not always calculated using the formula in USP <85>. Your firm does not ensure that each of your clients provides information regarding dosing of the drug product needed to calculate the MVD. For example, a. An endotoxin limit was not established for Clonidine/Ropivacaine (PF) Imcg/1mg/ml in 500mL 0.9% Sodium Chloride (injectable) submitted as sample #186092-01 by NECC and tested for endotoxins on 9/4/12. b. An endotoxin limit was not set for Baclofen PF (STOCK) 5000 meg/mL Injection submitted as sample #184445-01 by NECC and tested for endotoxins on 9/4/12. c. An endotoxin limit was not set established for CP2D submitted as sample #176189-01 by Ameridose and tested for endotoxins on 5/18/12.
- 4
No investigations for endotoxin OOS
4. Your firm has had 13 confirmed endotoxin failures for various drug products from October 2010-October 2012. There is no documentation of any investigations conducted into any endotoxin failures, including the failure of sample #186077-01 of Sodium Bicarbonate 150mEq/1000mL in Sterile Water for Injection that was submitted by NECC. SOP MBI-126 Microbiology Out-of-Specification Investigation (OOS), does not address investigation of OOS’s for endotoxin testing.
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.