Skip to content

ARL Biopharma, Inc.

840 Research Pkwy Ste 546, Oklahoma City, Oklahoma, United States

FEI 3003644883DUNS 169842007

Company: Analytical Research Laboratories · Linked by exact D-U-N-S match (169842007) in FDA registration data

FDA has 6 inspections on file for ARL Biopharma, Inc. in Oklahoma City, Oklahoma (FEI 3003644883), from 2010 to 2026. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

No warning letter, court action or import refusal is filed under this FEI.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for analysis, through Dec 31, 2027. Its registrant is ARL BioPharma, Inc. d.b.a. Analytical Research Laboratories (D-U-N-S 169842007).

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
6, 2010–2026
0 OAI · 2 VAI · 4 NAI
Inspections with 483 observations shown
1 of 6
1 posted 483 not matched to an inspection
Form 483s posted by FDA
↗
Requested under FOIA
1 request, latest Sep 15, 2022
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC

  1. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  3. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  4. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  5. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance4 citations published

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Incomplete USP <71> sterility practices

      1. Your firm states on the Microbiology Report that is issued to a client after sterility and/or fungal testing that the Test Method employed was USP <71>. However, your firm is not fully following all parts of USP <71> when performing sterility and/or fungal testing of human drug products. For example, a. USP <71> requires a Method Suitability Test be performed for all new products tested. Your firm does not have documentation to show that Method Suitability Testing has been performed for all drug products submitted for sterility testing by Ameridose, LLC and New England Compounding Center (NECC), both located in Massachusetts. For those drug products submitted by NECC and Ameridose, LLC, you have some documentation of bacteriostasis/fungistasis testing performed in 2006 & 2008 on a limited number of drug products, however there is no source documentation showing how the tests were performed, lot numbers of organisms or media used, and who performed the testing. b. USP <71> specifies the number of articles to be tested. While you provide reference to USP <71> for sample sizes, you do not ensure that your clients are submitting the required number of articles for testing. Most clients usually submit only [illegible] for sterility testing, including NECC and Ameridose.

    2. Observation 2

      Unvalidated potency (assay) methods

      2. Your firm has no documentation to show that all analytical methods used to test for potency (assay) have been validated for all drug products including drug products submitted for testing by NECC and Ameridose LLC. These include drug products such as Methylprednisolone Acetate, Heparin, Vasopressin, Triamcinolone Acetonide, and products containing Bupivacaine and Epinephrine. Analytical methods that are not validated and/ or not found in the USP that are used for potency testing of human drug products are not written, reviewed and approved.

    Show observations 3 to 4
    1. Observation 3

      Incomplete USP <85> endotoxin procedures

      3. Your firm states on the Microbiology Report that is issued to a client after endotoxin testing that the Test Method employed was USP <85>. Your firm is not fully following all parts of USP <85> when performing endotoxin testing of human drug products. Specifically, the Maximum Valid Dilution (MVD) is not always calculated using the formula in USP <85>. Your firm does not ensure that each of your clients provides information regarding dosing of the drug product needed to calculate the MVD. For example, a. An endotoxin limit was not established for Clonidine/Ropivacaine (PF) Imcg/1mg/ml in 500mL 0.9% Sodium Chloride (injectable) submitted as sample #186092-01 by NECC and tested for endotoxins on 9/4/12. b. An endotoxin limit was not set for Baclofen PF (STOCK) 5000 meg/mL Injection submitted as sample #184445-01 by NECC and tested for endotoxins on 9/4/12. c. An endotoxin limit was not set established for CP2D submitted as sample #176189-01 by Ameridose and tested for endotoxins on 5/18/12.

    2. Observation 4

      No investigations for endotoxin OOS

      4. Your firm has had 13 confirmed endotoxin failures for various drug products from October 2010-October 2012. There is no documentation of any investigations conducted into any endotoxin failures, including the failure of sample #186077-01 of Sodium Bicarbonate 150mEq/1000mL in Sterile Water for Injection that was submitted by NECC. SOP MBI-126 Microbiology Out-of-Specification Investigation (OOS), does not address investigation of OOS’s for endotoxin testing.

  6. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  7. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Sep 15, 2022. It asked for a Form 483.

  • Inspection date not given

    1 request. None of FDA’s log entries for these requests say records were released.

    1. Sep 15, 2022FDA-FOIA-2022-6630Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

FDA drug establishment registration

FDA registration search
Registered operations
Analysis
Registration runs through
Dec 31, 2027
Registrant
ARL BioPharma, Inc. d.b.a. Analytical Research LaboratoriesD-U-N-S 169842007

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003644883 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003644883. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.