Allegheny General Hospital Form 483, December 2015
FDA issued this Form 483 to Allegheny General Hospital in Pittsburgh, Pennsylvania (FEI 2521206) at the end of a device inspection that ended December 17, 2015. It lists 6 observations. FDA classified the inspection No Action Indicated (NAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Dec 17, 2015
- Devices · Postmarket Assurance: Devices
- FDA's outcome
- NAI
- No Action Indicated
- Posted by FDA
- Jan 17, 2024
- 2,953 days after the inspection
The 6 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Failure to report death (MDR)
OBSERVATION 1 The user facility did not submit FDA Form 3500A or electronic equivalent to FDA and the device manufacturer within ten working days after becoming aware of information that reasonably suggests that a device has or may have caused or contributed to the death of a patient of the facility. Specifically, an FDA Form 3500A was never submitted to the known device manufacturer or to the FDA after becoming aware of information that reasonably suggests that a Duodenoscope and/or a central line contributed to a patient death. The patient, with Medical Record [6(6)] underwent an ERCP procedure at another hospital and had positive CRB cultures at that facility. On [1~6l I], the patient was admitted and was found to have positive blood and sputum cultures upon admission. The patient underwent an ERCP procedure on b(6) J, and had repeated, subsequent positive CRE cultures following this procedure. On - negative blood cultures were taken, but on - positive CRE blood cultures were again found. The patient died on 1b(6) J This patient's "Infection Reporf", dated .....[f6t ) ] documents that the CRE infection found on - was a hospital and central-line associated infection.
- 2
Failure to report serious injury (MDR)
OBSERVATION 2 The user facility did not submit FDA Form 3500A or electronic equivalent to the known device manufacturer within 10 working days of becoming aware of information that reasonably suggests that a device has or may have caused or contributed to a serious injury to a patient of the facility. Specifically, an FDA Form 3500A was never submitted to the known device manufacturer or to the FDA after becoming aware of information that reasonably suggests that a Duodenoscope contributed to a serious injury to a patient of the facility. The patient, ) f with Medical Record (b(6) ) underwent an EGD procedure using a Duodenoscope on _l:>j~) .l and on - bile cultures were found to be positive for "ESCHERICHIACOLI FEWlire)unmLYeCarbapenemase producer 6 Isolate", as well as abdomen and aspirate cultures taken on (
- 3
MDR procedure not implemented
OBSERVATION3 Written MDR procedure(s) have not been implemented. Specifically, your "Safe Medical Device Act" medical device reporting procedure with Policy Manual No. 9018 has not been implemented. Furthermore, you have no documentation to show that all of the "Policy Guidelines" required per Policy Manual No. 9018 were followed or completed before or after you submitted Form FDA 3500 on 0{31 to the manufacturer concerning the adverse event with patient [1[5f3J b(6) and Medical Record [~~(3)b(6)]
- 4
Incomplete MDR event documentation
OBSERVATION4 MDR event files do not contain or reference all adverse event information in the possession of the reporting entity, including documentation of the deliberations and decision making process used to determine if an event was or was not reportable. Specifically, you have no documentation of the deliberations and decision making process used to determine if the death of E.M.F., with patient identifier 571824 and Medical Record #00912139, was or was not reportable.
- 5
Incomplete information on MDR report
OBSERVATION5 The user facility report submitted on FDA Form 3500A did not include all information reasonably known. Specifically, a) on f6\ J you submitted FDA Form 3500 for voluntary reporting, to the medical device manufacturer due to a serious injury patient, instead of the required FDA Form 3500A for [illegible]. The patient b(6) with patient Identifier b( ) b( ) d Medical Record # b(3 b(6) was admitted on 15(3) 15(6) and underwent an b(3) using ab(3) on the same d~ days later on /b( ) The blood cultures were drawn which were found positive for '-6(~~) Additional positive blood cultures were also drawn on subsequent -f\ days. On . blood cultures (re-drawn which were found to be negative for b(3) and the patient was discharged the next day on ( ) No bile or sputum cultures were taken. The patient died on an [illegible] date in 15(6) later at another facility, and your 3)' l b(3) b) The Form FDA3500 you submitted to the medical device manufacturer on ~l concerning patient 115(3) b(6) and Medical Record f6{31616Ti did not include all required information that was reasonably known to you at that time. The following information was reasonably known to you, but was not submitted at that time: patient weight, patient followup or required treatment, brand name, product code, model number, lot number or other identifying number of the associated medical device.
- 6
No timely MDR transmission system
OBSERVATION 6 The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA and manufacturers. Specifically, your "Safe Medical Device Act" medical device reporting procedure with Policy Manual No. 9018 does not specify that reports of death must be submitted to the FDA and the manufacturer, if known, and that reports of serious injury must be submitted to the manufacturer.
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.