Skip to content

Allegheny General Hospital

320 E North Ave, Pittsburgh, Pennsylvania, United States

FEI 2521206

FDA has one inspection on file for Allegheny General Hospital in Pittsburgh, Pennsylvania (FEI 2521206). It ended on and was classified NAI, no action indicated. FDA has also posted a Form 483 dated Oct 25, 2016 that FDAgent has not matched to an inspection on the Data Dashboard.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA posted the Form 483 from that inspection. FDAgent read its 6 observations from the PDF; they are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
1, in 2015
Inspections with 483 observations shown
1 of 1
1 posted 483 not matched to an inspection
Form 483s posted by FDA
↗
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC

  1. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Failure to report death (MDR)

      OBSERVATION 1 The user facility did not submit FDA Form 3500A or electronic equivalent to FDA and the device manufacturer within ten working days after becoming aware of information that reasonably suggests that a device has or may have caused or contributed to the death of a patient of the facility. Specifically, an FDA Form 3500A was never submitted to the known device manufacturer or to the FDA after becoming aware of information that reasonably suggests that a Duodenoscope and/or a central line contributed to a patient death. The patient, with Medical Record [6(6)] underwent an ERCP procedure at another hospital and had positive CRB cultures at that facility. On [1~6l I], the patient was admitted and was found to have positive blood and sputum cultures upon admission. The patient underwent an ERCP procedure on b(6) J, and had repeated, subsequent positive CRE cultures following this procedure. On - negative blood cultures were taken, but on - positive CRE blood cultures were again found. The patient died on 1b(6) J This patient's "Infection Reporf", dated .....[f6t ) ] documents that the CRE infection found on - was a hospital and central-line associated infection.

    2. Observation 2

      Failure to report serious injury (MDR)

      OBSERVATION 2 The user facility did not submit FDA Form 3500A or electronic equivalent to the known device manufacturer within 10 working days of becoming aware of information that reasonably suggests that a device has or may have caused or contributed to a serious injury to a patient of the facility. Specifically, an FDA Form 3500A was never submitted to the known device manufacturer or to the FDA after becoming aware of information that reasonably suggests that a Duodenoscope contributed to a serious injury to a patient of the facility. The patient, ) f with Medical Record (b(6) ) underwent an EGD procedure using a Duodenoscope on _l:>j~) .l and on - bile cultures were found to be positive for "ESCHERICHIACOLI FEWlire)unmLYeCarbapenemase producer 6 Isolate", as well as abdomen and aspirate cultures taken on (

    Show observations 3 to 6
    1. Observation 3

      MDR procedure not implemented

      OBSERVATION3 Written MDR procedure(s) have not been implemented. Specifically, your "Safe Medical Device Act" medical device reporting procedure with Policy Manual No. 9018 has not been implemented. Furthermore, you have no documentation to show that all of the "Policy Guidelines" required per Policy Manual No. 9018 were followed or completed before or after you submitted Form FDA 3500 on 0{31 to the manufacturer concerning the adverse event with patient [1[5f3J b(6) and Medical Record [~~(3)b(6)]

    2. Observation 4

      Incomplete MDR event documentation

      OBSERVATION4 MDR event files do not contain or reference all adverse event information in the possession of the reporting entity, including documentation of the deliberations and decision making process used to determine if an event was or was not reportable. Specifically, you have no documentation of the deliberations and decision making process used to determine if the death of E.M.F., with patient identifier 571824 and Medical Record #00912139, was or was not reportable.

    3. Observation 5

      Incomplete information on MDR report

      OBSERVATION5 The user facility report submitted on FDA Form 3500A did not include all information reasonably known. Specifically, a) on f6\ J you submitted FDA Form 3500 for voluntary reporting, to the medical device manufacturer due to a serious injury patient, instead of the required FDA Form 3500A for [illegible]. The patient b(6) with patient Identifier b( ) b( ) d Medical Record # b(3 b(6) was admitted on 15(3) 15(6) and underwent an b(3) using ab(3) on the same d~ days later on /b( ) The blood cultures were drawn which were found positive for '-6(~~) Additional positive blood cultures were also drawn on subsequent -f\ days. On . blood cultures (re-drawn which were found to be negative for b(3) and the patient was discharged the next day on ( ) No bile or sputum cultures were taken. The patient died on an [illegible] date in 15(6) later at another facility, and your 3)' l b(3) b) The Form FDA3500 you submitted to the medical device manufacturer on ~l concerning patient 115(3) b(6) and Medical Record f6{31616Ti did not include all required information that was reasonably known to you at that time. The following information was reasonably known to you, but was not submitted at that time: patient weight, patient followup or required treatment, brand name, product code, model number, lot number or other identifying number of the associated medical device.

    4. Observation 6

      No timely MDR transmission system

      OBSERVATION 6 The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA and manufacturers. Specifically, your "Safe Medical Device Act" medical device reporting procedure with Policy Manual No. 9018 does not specify that reports of death must be submitted to the FDA and the manufacturer, if known, and that reports of serious injury must be submitted to the manufacturer.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2521206 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2521206. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.