Skip to content

Cooler Heads Care, Inc.

7197 Convoy Ct Ste 5, San Diego, California, United States

FEI 3021619527

FDA has 2 inspections on file for Cooler Heads Care, Inc. in San Diego, California (FEI 3021619527), from 2024 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
2, 2024–2026
0 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest Sep 10, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 8.3.1

      Nonconforming product not adequately evaluated

      Nonconforming product was not evaluated to determine the need for an investigation and notification of any external party responsible for the nonconformity.

    2. ISO 13485:2016 Clause 7.4.1

      Records of supplier (re)evaluation, selection, and monitoring not maintained

      Records of the results of evaluation, selection and monitoring of supplier capability or performance were not maintained.

    3. ISO 13485:2016 Clause 8.2.1

      Information gathered in feedback process not used as input into risk management

      Information gathered in the feedback process does not serve as potential input into risk management for monitoring and maintaining product requirements as well as the product realization or improvement processes.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Sep 10, 2026. Each one asked for a Form 483.

  • Inspection ending Aug 6, 2026VAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 22, 2026. FDAgent holds no copy that FDA posted.

    1. Sep 10, 2026FDA-FOIA-2026-8793Form 483Records released
    2. Sep 1, 2026FDA-FOIA-2026-8487Form 483· 1 of 5 sitesRecords released
  • Inspection ending Apr 5, 2024VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Aug 30, 2024FDA-FOIA-2024-7645Form 483· 1 of 2 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 1 section of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3021619527 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3021619527. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.