Biocon Biologics Limited
Block No. M1, M2 and M6, Q1 (QC3 and QC10) and W3, 20th KM Hosur Road, Electronics City, Bengaluru, India
FEI 3015283245DUNS 675486243
Company: Biocon Biologics Limited · Linked by exact D-U-N-S match (860350848) in FDA registration data
FDA has 2 inspections on file for Biocon Biologics Limited in Bengaluru, India (FEI 3015283245), from 2024 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA posted the Form 483 from that inspection. FDAgent read its 5 observations from the PDF; they are listed below.
3 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, API manufacturing, packing and analysis, through Dec 31, 2027. Its registrant, Biocon Biologics Limited (D-U-N-S 860350848), has registered 2 other sites.
Sources:FDA firm profile ↗Form 483 PDF ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2024–2025
- 0 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 1 of 2
- 1 posted 483 not matched to an inspection
- Requested under FOIA
- 2 requests, latest Nov 22, 2024
- Warning letters
- None on file
- Import refusals
- 3 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 5 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Inadequate sanitation and standing water
Written sanitation and facility maintenance procedures are not established or followed. Specifically, A) The process facility maintenance and cleanliness is inadequate for Grade C and D manufacturing areas. This includes the following deficiencies noted during the facility walkthrough on August 26th, 2025: 1. (b)(4) standing puddle within the Grade C room and adjacent to the [illegible]. 2. A soiled mop labeled as "Grade D Area Cleaning" within the Grade C room that is adjacent to the Grade C batches. 3. (b)(4) standing puddle underneath the (b)(4) area. 4. Liquid standing puddle underneath (b)(4) tank within (b)(4) area. 5. Large liquid standing puddle underneath (b)(4) tank within (b)(4) area. 6. A large standing puddle could on the facility floor of (b)(4) area in-process (b)(4) (b)(4) tank (b)(4) according to the Deputy Manager of Production. The (b)(4) standing puddle of liquid is likely (b)(4). 7. Standing puddle underneath (b)(4) tank within (b)(4) area. 8. (b)(4) standing puddle underneath (b)(4) tank within (b)(4) area. 9. Liquid standing puddle underneath (b)(4) tank within (b)(4) area. 10. The esfracting units within the in-process area, including advisible stains on the pleated filter surfaces. B) The process facility maintenance is inadequate. This includes the following deficiencies noted during the facility walkthrough on August 26th, 2025: The (b)(4) used for (b)(4) manufacturing operations was leaking consistently with a container bucket placed directly underneath the leak. However, there was no written notification or investigation into the leak until the current inspection.
- Observation 2
Equipment cleaning and maintenance lapses
Written procedures are not established or followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug substance. Specifically, On August 29th, 2025, I observed visible rust spots within the Grade B laminar Air Flow (LAF) and visible rust/removable material on the outer LAF surfaces of the Grade C environment. The LAF is used for activities for (b)(4) bulk drug substance batches with a bioburden specification of NMT (b)(4) FU/mg of protein and bacterial endotoxins specification of NMT (b)(4) FU/mg of protein. This included: 1. Visible rust surrounding the seams and ridges of the valve and knob stationed within the Grade B LAF. 2. Visible rust on the screw part within the Grade B side of the LAF view screen. 3. Sealant observed throughout the corner and edges of the Grade B LAF. 4. Visible rust observed within the edges of Grade B LAF. 5. Visible rust and brown material that was removable with a wipe on the outer surface edges on the Grade B LAF housed in the Grade C environment. Dynamic Air flow visualization pattern studies for the Grade B Laminar Air Flow (LAF) did not include the following equipment and materials routinely used during activities for (b) bulk drug substance batches with a bioburden specification of NMT (b)(4) FU/mg of protein and bacterial endotoxin specification of NMT (b)(4) FU/mg of protein: 1. (b)(4) that is kept inside the Grade B LAF during (b)(4) activities. 2. Other required materials loaded within the Grade B (e.g., [illegible]). In addition, Requalification of HAF01, Document No. RQ-1-0007.01, Dated: July 10, 2024, listed the following demo simulations as executed, completed, and signed off with the following Test Status: PASS Actual Result: Unidirectional Sweeping action observed. However, there was no airflow study evaluation of the (b)(4) used during activities. In addition, there is no written evaluation of the eddy currents observed to be bouncing in the LAF benchtop at the following dynamic smoke video timepoints: On August 26th, 2025, I observed visible liquid droplets throughout the view glass within the (b)(4) labeled as Cleaned and Ready for Use on the same day. The level of droplets obstructed the view within the (b)(4) (this includes the following equipment checklist drying steps marked as completed: Equipment Drying: - (b)(4)). According to the Global Compliance Head, the presence of liquid water droplets within equipment is acceptable post cleaning and drying operations. However, the acceptance level for the varying level of equipment dryness and water retention prior to equipment release is not defined or specified within the equipment release checklist. (b)(4) used to store (b)(4) during (b)(4) operations within (b)(4) area was observed to be visibly discolored. This is despite SOP Document No. (b)(4) RLT.0:~0 P:011 2 Procedure for Identification and Usage of (b)(4) Effective Date: July 31, 2025. Section 6.2 d) Physical verification of (b)(4) for any discoloration, deformation, elongation etc. There is no visual challenge performed for operator check for physical verification of (b)(4). E) (b)(4) labeled as cleaned and ready for use on August 25, 2025 had visible wear, metallic scratches and dents throughout rim of the (b)(4) used to the (b)(4) process stage.
Show observations 3 to 5
- Observation 3
Insufficient OOS investigation and review
There is a failure to thoroughly review any unexplained discrepancy and make a record of the conclusions and follow-up. A) Out of Specification (OOS) Investigation Report No. 199753 was initiated on June 10, 2024, due to OOS observed result: (b)(4) FU/mL (Limit NMT (b)(4) FU/mL) for (b)(4) sampling location description: (b)(4). According to the report, from the identified isolates (Ralstonia insidiosa, Elizabethkingia miricola, Chryseobacterium indologenes) most probable cause could be current sanitization procedure not able to eliminate these isolates thus that could lead to such localized microbial load build up. Additionally, Due to the rust on the surface on the devices might have probably led to potential ingress of the microbes on the (b)(4) which might have carried into the collected sample resulting into OOS for the sampling points. In addition, CAPA Number: 1,21318 was initiated as an additional control to reduce microbial ingress. 192406 is logged to revise the SOP (ii) OP/0158 instruction to clean the hoses used for (b)(4) collection with (b)(4) from the last usage. However, OOS Investigation Report No. 199753, Date: June 10, 2024 did not include a detailed assessment and retrospective review of previous similar OOSs or events, including: OOS Report No. 186956 initiated on April 8, 2024 due to OOS result (b)(4) FU/mL (Limit NMT (b)(4) FU/mL) observed for Sampling point: (b)(4), Sampling location description: (b)(4). The investigation also states that from the identified isolates (Ralstonia insidiosa, Kocuria palustris) the most probable cause could be current sanitization procedure not able to eliminate these isolates thus that could lead to such localized microbial load build up. CAPA Number 11318 was initiated as an additional control to reduce microbial ingress. 192406 is logged to revise SOP/0158 instruction to clean the hoses used for (b)(4) collection with (b)(4) from the last usage.
- Observation 4
Quality unit procedures not followed
The responsibilities and procedures applicable to the quality control unit are not in writing or fully followed. Specifically, A) Notification No: 1000387247, Leak Test Failed, Dated: January 24, 2024 was raised due to leak test being conducted more times and having multiple failed results. The (b)(4) of the complete system found that leak was of (b)(4) side. A material used for engineering grade repair system for repairing and rebuilding machinery and equipment was applied on the (b)(4) with the failed leak test. However, there is no detailed impact assessment, risk analysis, cross-functional approval, and post implementation verification for the effectiveness of the equipment repair operations. B) During the inspection walkthrough on August 26, 2025, I observed a crumbled up sterilization label for (b)(4) directly on top of the (b)(4) container in Grade C (b)(4) loading area with the following hand written details: Equipment Name: Sterilized: 24/08/2025 Use Before: Sign and Date: 24/08/2025. However, there is no documented reasoning as to why the label was crumbled up.
- Observation 5
Laboratory controls and equipment follow-up lacking
Laboratory Control Mechanisms are not established. A) On September 01, 2025, I observed that there is no final disposition record for one (1) out of 1 (b)(4) media plates Expired On: August 31, 2025 labeled as Environmental Monitoring (EM) sampling activities for (b)(4) on August 12, 2025, AR No. (b)(4) MICRO-25-005 within the laboratory information management software. B) There is no notification or documented follow-up to multiple errors observed within QC Analytical Balances. This consists of Analytical Balance NTP Server Error: "Because of the server down/network failure temporary." C) The following notification reader errors were displayed on the BET Kinetic software after being released for completion of Performance Qualification (PQ) on August 27, 2024: Error #0400 code at: (b)(4). According to the software manual: Error #0400: Carrier x-axis failed position verify. Error #0401: Filter wheel motor failed position verify. Errors 0400 and 0401 indicate that an axis failed its Position Verify test. After moving a predefined number of steps from a home position, the motor should return to the home position in the proper amount of time and steps. If the axis moves back to its home position in the wrong amount of time or in too few or too many steps, the test fails. Probable causes: o Mechanical obstruction o Carrier belt or pulley slipping o Loose filter wheel hub o Loose filter may be impeding filter wheel movement o Defective motor drive circuit or 12-volt power problem. However, there was no further assessment or documented follow-up with the equipment vendor, engineering team, and performance qualification status.
- VAIObjectionable conditions found; FDA took no regulatory actionVeterinary · Pre-Approval Eval. of Animal Drugs and Food AdditivesCitations not published
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Nov 22, 2024.
Inspection ending Feb 28, 2024VAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Nov 22, 2024FDA-FOIA-2024-10342EIRWithdrawn
- Aug 12, 2024FDA-FOIA-2024-7163Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3015283245, one row per refused product line.
3 refusals from 2020-09-25 to 2025-09-12 · charged UNAPPROVED on 2, NOT LISTED on 1 · no FDA sample analysis
FDA firm profile- 2025-09-12
SIROLIMUS (IMMUNOSUPPRESSIVE)64TCE12
UNAPPROVED 505(a), 801(a)(3); UNAPPROVED NEW DRUG
Division of Southeast ImportsAEK-7475392-0/30002/1
- 2022-06-10
SIROLIMUS (IMMUNOSUPPRESSIVE)64TCS12
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of West Coast ImportsB4S-0008135-4/11/3
- 2020-09-25
CLOBETASOL PROPIONATE (GLUCOCORTICOID)64LAO67
UNAPPROVED 505(a), 801(a)(3); UNAPPROVED NEW DRUG
Division of Southeast ImportsSCS-1762084-2/11/1
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing, API manufacturing, packing, analysis
- Registration runs through
- Dec 31, 2027
- Registrant
- Biocon Biologics LimitedD-U-N-S 860350848
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Other sites with the same registrant
FDA’s registration file names Biocon Biologics Limited (D-U-N-S 860350848) as the registrant for this site and the 2 others below. That exact D-U-N-S number is the only link. A registrant is the firm that filed the registration, which is not always the owner.
- BIOCON SDN.BHD.Iskandar Puteri, Malaysia · FEI 30112482481 inspection1latestSep 27, 2024VAINone on file
- Biocon Biologics LimitedBengaluru, India · FEI 300398147511 inspections11latestJul 26, 2024VAINone on file
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. This site’s records are filed under FEI 3015283245 in FDA’s data; other sites under its registrant are listed in summary only. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every inspection and enforcement record on this page is filed under FEI 3015283245. Other sites filed under the same registrant D-U-N-S are listed in summary only. Records filed under another FEI or name are otherwise not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.