Ionetix Corporation

1950 S Tamiami Trl, Sarasota, Florida, United States

FEI 3013422537DUNS 080610313

Company: Ionetix Corporation · Linked by exact D-U-N-S match (967706008) in FDA registration data

FDA has 2 inspections on file for Ionetix Corporation in Sarasota, Florida (FEI 3013422537), from 2019 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA posted the Form 483 from that inspection. FDA published 4 citations from this inspection, all under 21 CFR Part 212, the manufacturing rules for PET drugs. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for PET drug production, through Dec 31, 2027. Its registrant, Ionetix Corporation (D-U-N-S 967706008), has registered 8 other sites.

Sources:FDA firm profile ↗Form 483 PDF ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
2, 2019–2026
0 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
1 of 2
Form 483s posted by FDA
↗
Requested under FOIA
2 requests, latest Jun 8, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 4 citations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Written control procedures

      You did not have written production and process control procedures to ensure that any deviations from the procedures are justified.

    2. Equipment procedures

      You did not implement procedures and document your activities in accordance with your procedures to ensure that all equipment is suitable for its intended purposes and properly installed, maintained, and capable of repeatedly producing valid results that could reasonably be expected to adversely affect the identity, strength, quality, or purity of a PET drug, or give erroneous or invalid test results when improperly used or maintained.

    3. Specifications and processes

      You did not approve or reject, before implementation, specifications to ensure that they maintain the identity, strength, quality, and purity of a PET drug.

    4. Use of reliable suppliers

      You did not use a reliable supplier as a source of each lot of component.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Jun 8, 2026. Each one asked for a Form 483.

  • Inspection ending Apr 2, 2026VAI

    1 request. In the inspection history above. FDA has posted its Form 483.

    1. Jun 8, 2026FDA-FOIA-2026-5622Form 483· 1 of 5 sitesRecords released
  • Inspection date not given

    1 request. None of FDA’s log entries for these requests say records were released.

    1. Jul 22, 2025FDA-FOIA-2025-8936Form 483 + EIR· 1 of 4 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 4 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.

  • 21 CFR 212.20What activities must I perform to ensure drug quality?1 of 1
  • 21 CFR 212.30What requirements must my facilities and equipment meet?1 of 1
  • 21 CFR 212.40How must I control the components I use to produce PET drugs and the containers and closures I package them in?1 of 1
  • 21 CFR 212.50What production and process controls must I have?1 of 1

FDA drug establishment registration

FDA registration search
Registered operations
PET drug production
Registration runs through
Dec 31, 2027
Registrant
Ionetix CorporationD-U-N-S 967706008

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Other sites with the same registrant

FDA’s registration file names Ionetix Corporation (D-U-N-S 967706008) as the registrant for this site and the 8 others below. That exact D-U-N-S number is the only link. A registrant is the firm that filed the registration, which is not always the owner.

  • Ionetix CorporationJenkintown, Pennsylvania · FEI 30152877912 inspections2latestJul 23, 2026VAINone on file
  • Ionetix CorporationKansas City, Kansas · FEI 30152658242 inspections2latestOct 31, 2025VAINone on file
  • Ionetix CorporationMiami Beach, Florida · FEI 30221149001 inspection1latestSep 26, 2025VAINone on file
  • Ionetix CorporationMiami, Florida · FEI 30179538151 inspection1latestMay 1, 2025VAINone on file
  • IonetixPittsburgh, Pennsylvania · FEI 30164608011 inspection1latestJan 19, 2024VAINone on file
  • Ionetix CorporationEl Paso, Texas · FEI 30164586431 inspection1latestSep 14, 2023VAINone on file
  • Ionetix CorporationKnoxville, Tennessee · FEI 3025767691Registered; no FDA inspection on file
  • Ionetix CorporationMiami, Florida · FEI 3030286261Registered; no FDA inspection on file

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. This site’s records are filed under FEI 3013422537 in FDA’s data; other sites under its registrant are listed in summary only. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every inspection and enforcement record on this page is filed under FEI 3013422537. Other sites filed under the same registrant D-U-N-S are listed in summary only. Records filed under another FEI or name are otherwise not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.