Skip to content

Pattanam D. Srinivasan, MD

Advanced Interventional Pain Ctr, 613 W Lincoln Rd Ste A, Kokomo, Indiana, United States

FEI 3010277574

FDA has one inspection on file for Pattanam D. Srinivasan, MD in Kokomo, Indiana (FEI 3010277574). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 28 citations from this inspection. They are listed below.

No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2013
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring28 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Record retention inadequate

      Required records were not all maintained during the investigation, for a period of two years after the date on which an investigation was terminated or completed, and for a period of two years after the date that the records were no longer required for purposes of supporting a premarket approval application or a notice of completion of a product development protocol.

    2. Copy of consent form not provided

      A copy of the written consent form which had been approved by the IRB and signed and dated by the subject or the subject's legally authorized representative, was not provided to the subject or the subject's legally authorized representative at the time of consent.

    3. Investigator device accountability inadequate

      Records of receipt and use of a device that relate to the type and quantity, dates of receipt, and batch number or code mark of the device are not all complete.

    4. Number of subjects in study

      The informed consent document did not state the approximate number of subjects involved in the study.

    5. Promotion or test marketing

      An investigational device was promoted and test marketed prior to FDA approval/clearance.

    Show 23 more citations
    1. Compensation, medical treatment in event of injury

      For research involving more than minimal risk, the informed consent document lacked an explanation as to whether any compensation and medical treatments were available if injury occurred, and, if so, of what they consisted and where further information might be obtained.

    2. Needs IDE; diagnostic device exemption conditions not met

      An investigation requires an IDE, in that the device is a diagnostic device which is invasive, by design or intention introduces energy into a subject, and is used as a diagnostic procedure without confirmation of the diagnosis by another, medically established diagnostic product or procedure.

    3. Participation; refusal and discontinuance

      The informed consent document did not contain a statement that participation was voluntary, refusal to participate would involve no penalty or loss of benefits to which the subject was otherwise entitled, and the subject might discontinue participation at any time without penalty or loss of benefits to which the subject was otherwise entitled.

    4. No investigation agreement statement of commitment

      A signed agreement that includes a statement of the investigator's commitment to conduct an investigation in accordance with the agreement, investigational plan, applicable FDA regulations, conditions of approval imposed by the reviewing IRB, and conditions of approval imposed by the FDA was not obtained from each participating investigator.

    5. Investigator records of relevant observations inadequate

      Records for each subject concerning condition during the course of the investigation are not all complete.

    6. Exculpatory language, waiving of rights

      The general requirements for informed consent were not met in that the consent form contained exculpatory language through which the subject or the subject's representative was made to waive or appear to waive some of the subject's legal rights and that released or appeared to release the investigator, sponsor, the institution, or its agents from liability for negligence.

    7. Possible additional costs to the subject

      The informed consent document did not include a statement of any additional costs to the subject that might result from participation in the research, as appropriate.

    8. Procedures, identification of those which were experimental

      The informed consent document did not contain a description of the procedures to be followed and identification of any procedures which were experimental.

    9. Investigator final report

      A final report to the sponsor and reviewing IRB was not submitted within three months after completion of the investigation.

    10. Consequences of withdrawing

      The informed consent document did not provide information regarding the consequences of a subject's decision to withdraw from the research, and procedures for orderly termination of participation by the subject.

    11. Significant new findings

      The informed consent document did not include a statement that significant new findings developed during the course of the research, which might relate to the subject's willingness to continue participation, would be provided to the subject.

    12. Termination by investigator

      The informed consent document did not include a statement of anticipated circumstances under which the subject's participation might be terminated by the investigator, without regard to the subject's consent.

    13. Consent form not approved/signed/dated

      Informed consent was not properly documented in that the written informed consent used in the study was not approved by the IRB.

    14. No FDA approval for treatment IDE

      A treatment IDE did not have FDA approval.

    15. Alternate procedures, courses of treatment

      There was no disclosure in the informed consent document of appropriate alternate procedures or courses of treatment, if any, that might be advantageous to the subject.

    16. Statement of research, purpose, duration of participation

      The informed consent document did not contain a statement that the study involved research and an explanation of the purposes of the research, and the expected duration of the subject's participation.

    17. Whom to contact

      The informed consent document lacked an explanation of whom to contact for answers to pertinent questions about the research and research subjects' rights and in the event of a research-related injury to the subject.

    18. Reasonably foreseeable risks or discomforts

      The informed consent document lacked a description of reasonably foreseeable risks or discomforts to the subject.

    19. No IRB approval for treatment use

      A treatment use study of an investigational device did not have IRB approval.

    20. Investigator non-compliance with agreement/plan/regulations

      An investigation was not conducted according to the investigational plan and applicable FDA regulations.

    21. Investigator's subject records inadequate

      Records of each subject's case history and exposure to the investigational device are not all accurate.

    22. Confidentiality, FDA inspection of records

      There was no statement in the informed consent document that described the extent, if any, to which confidentiality of records identifying the subject would be maintained and noted the possibility that the Food and Drug Administration might inspect the records.

    23. Statement of risks

      The informed consent document did not contain a statement that the test article or procedure might involve risks to the subject (or to the embryo or fetus, if the subject is or may become pregnant) that are currently unforeseeable.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3010277574 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3010277574. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.