Skip to content

Steribite, LLC

1877 Eagle Pass, Wooster, Ohio, United States

FEI 3009662229

FDA has 2 inspections on file for Steribite, LLC in Wooster, Ohio (FEI 3009662229), from 2017 to 2026. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 12 citations from this inspection. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2017–2026
1 OAI · 0 VAI · 1 NAI
Inspections with 483 observations shown
1 of 2
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest Sep 15, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.3.1

      Design and development procedures not documented

      Design and development procedures were not documented.

    2. ISO 13485:2016 Clause 7.5.1

      Production controls not adequate

      Production controls do not include documentation of procedures and methods for the control of production, implementation of monitoring and measurement of process parameters and product characteristics and implementation of product release, delivery, and post-delivery activities.

    3. Devices subject to device identification GUDID data submission requirements.

      The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.

    4. ISO 13485:2016 Clause 4.1.2

      Risk-based approach not applied to QMS processes

      A risk-based approach to the control of the appropriate processes needed for the quality management system has not been applied.

    5. ISO 13485:2016 Clause 7.5.6

      Planned results not consistently met during validation of processes

      Validation of processes for production and service provision did not demonstrate the ability of the processes to achieve planned results consistently.

    Show 7 more citations
    1. ISO 13485:2016 Clause 7.5.6

      Computer software used in production and service provision not validated

      Software used in production and service provision was not validated prior to initial use and after changes to such software or its application.

    2. ISO 13485:2016 Clause 5.6.1

      QMS not reviewed by top management at documented planned intervals

      Top management did not review the quality management system at documented planned intervals to ensure its continued suitability, adequacy, and effectiveness.

    3. ISO 13485:2016 Clause 6.2(e)

      Records of personnel competence not maintained

      Appropriate records of personnel training were not maintained.

    4. ISO 13485:2016 Clause 4.1.5

      External party controls do not include written quality agreements

      Control over outsourced processes that affect product conformity to requirements do not include written quality agreements.

    5. ISO 13485:2016 Clause 7.1

      Processes needed for product realization not planned and developed

      Processes needed for product realization were not planned and developed.

    6. ISO 13485:2016 Clause 8.3.1

      Procedure for nonconforming product not documented

      Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not adequately documented.

    7. ISO 13485:2016 Clause 4.2.3

      Medical device file not established and maintained

      Medical device files have not been established and maintained.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Sep 15, 2026. Each one asked for a Form 483.

  • Inspection ending Apr 8, 2026OAI

    3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 10, 2026. FDAgent holds no copy that FDA posted.

    1. Sep 15, 2026FDA-FOIA-2026-8918Form 483· 1 of 5 sitesRecords released
    2. Jul 16, 2026FDA-FOIA-2026-6784Form 483Records released
    3. Jun 24, 2026FDA-FOIA-2026-6048Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 1 section of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.

  • 21 CFR 830.300Devices subject to device identification data submission requirements1 of 1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3009662229 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3009662229. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.