Incumedx LLC
46790 Lakeview Blvd, Fremont, California, United States
FEI 3009606113
FDA has one inspection on file for Incumedx LLC in Fremont, California (FEI 3009606113). It ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 8 citations from this inspection, all under 21 CFR Part 58, the good laboratory practice rules for nonclinical studies. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 1, in 2014
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Final report: description of methods
The final study report did not include a description of the methods used.
Protocol: experimental design
Not all protocols contained a description of the experimental design, including the methods for the control of bias.
Final report: name, address, dates
The final study report did not include the name and address of the facility performing the study and the dates on which the study was initiated and completed.
Protocol: date and signature
Not all protocols contained the date of approval of the protocol by the sponsor and the dated signature of the study director.
Archives: data, documentation, specimens
Not all raw data generated as a result of a nonclinical laboratory study were retained.
Show 3 more citations
Protocol: sponsor name and address
Not all protocols contained the name of the sponsor and the name and address of the testing facility at which the study is being conducted.
Final report: not signed by study director
Not all final reports were signed and dated by the study director.
Notifying contractor of GLP status
Not all consulting laboratories, contractors, or grantees were notified that the study must be conducted in compliance with FDA GLP regulations.
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 58.10Applicability to studies performed under grants and contracts1 of 1
- 21 CFR 58.120Protocol1 of 1
- 21 CFR 58.185Reporting of nonclinical laboratory study results1 of 1
- 21 CFR 58.190Storage and retrieval of records and data1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3009606113 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3009606113. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.