Skip to content

Rodo Medical Inc

2635 N 1st St Ste 218, San Jose, California, United States

FEI 3009113203

FDA has one inspection on file for Rodo Medical Inc in San Jose, California (FEI 3009113203). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 16 citations from this inspection. They are listed below.

No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2014
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring16 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Sponsor device disposition records inadequate

      Records of disposition of a device which describe the batch number or code marks of any devices disposed of by the investigator or another person and the reasons for and method of disposition reports are not all complete.

    2. No curriculum vitae in investigator agreement

      The signed investigator agreements obtained from each participating investigator do not all include the investigator's curriculum vitae.

    3. Sponsor records re QSR/GMP & non-significant risk studies

      For an investigation subject to the abbreviated requirements, records that include a statement of the extent to which the regulations covering current good manufacturing practice would be followed in manufacturing the device reports are not all complete.

    4. Procedures, identification of those which were experimental

      The informed consent document did not contain identification of any procedures which were experimental.

    5. Sponsor provide device information to investigators

      Not all investigators participating in the investigation were supplied with copies of the investigational plan.

    Show 11 more citations
    1. Sponsor correspondence records inadequate

      Records relating to correspondence with an investigator and an IRB, including required reports are not all complete.

    2. No evaluation of unanticipated adverse effects

      An evaluation of reported unanticipated adverse device effects was not conducted.

    3. No financial disclosure info in investigator agreement

      A signed agreement was not obtained from each participating investigator that includes sufficient accurate financial disclosure information to allow the sponsor to submit a complete and accurate certification or disclosure statement and a commitment to promptly update financial disclosure information if any relevant changes occur during the investigation and for one year following completion of the study.

    4. Monitors not qualified

      Monitors that are qualified by training and experience were not chosen to monitor the investigational study.

    5. Sponsor records of adverse device effects & complaints

      Records concerning complaint reports are not all complete.

    6. Sponsor records of investigator agreements inadequate

      Signed investigator agreements that include the required financial disclosure information are not all complete.

    7. Sponsor records re device use in non-significant risk study

      For an investigation subject to the abbreviated requirements, records which include the objectives of the investigation are not all complete.

    8. Sponsor began study before IRB/FDA approval

      An investigation was initiated before FDA approval.

    9. Record retention inadequate

      Required records were not all maintained for a period of two years after the date that the records were no longer required for purposes of supporting a premarket approval application or a notice of completion of a product development protocol.

    10. No info on relevant experience in investigator agreement

      A signed agreement that includes the specifics of the investigator's relevant experience was not obtained from each participating investigator.

    11. Investigator(s) not qualified

      One or more investigators were selected who were not qualified by training to investigate a device.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3009113203 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3009113203. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.