Skip to content

John Meadors

12458 Timberline Dr, Highland, Utah, United States

FEI 3009093259

FDA has one inspection on file for John Meadors in Highland, Utah (FEI 3009093259). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 8 citations from this inspection. They are listed below.

No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2018
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigator non-compliance with agreement/plan/regulations

      An investigation was not conducted in accordance with the investigational plan and applicable FDA regulations.

    2. Investigator records of relevant observations inadequate

      Records for each subject concerning previous medical history, condition upon entering the investigation, condition during the course of the investigation and all diagnostic test results are not all complete.

    3. Investigator records of exposure to device inadequate

      Records of each subject's exposure to the device, including the date and time of each use and the use of any other therapy are not all complete.

    4. Notice of transfer of responsibility for records

      A notice of transfer of responsibility for the maintenance of required records was not submitted.

    5. Investigator records of informed consent inadequate

      Records documenting that informed consent was obtained for each subject prior to participation in the study are not all complete.

    Show 3 more citations
    1. Subject participation prior to study approval

      Subjects were allowed to participate in an investigation prior to obtaining IRB and FDA approval to conduct the investigation.

    2. Investigator's subject records inadequate

      Records of each subject's case history and exposure to the investigational device are not all complete.

    3. Investigator device accountability inadequate

      Records of receipt, use and disposal of a device that relate to the type and quantity, dates of receipt and batch number or code mark of the device are not all complete.

Regulations FDA cited here

The 2 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3009093259 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3009093259. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.