Skip to content

21 CFR 812.110: Specific responsibilities of investigators

FDA's inspection data lists 365 Form 483 citations of 21 CFR 812.110 at 341 facilities, from October 2008 to September 14, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
365
At 348 inspections
Facilities cited
341
Distinct FEIs
Most recent citation
Sep 14, 2026
Roderick H. Tung, M.D.
Warning letters
0
State a violation of it

The regulation

The current text of 21 CFR 812.110 from the Electronic Code of Federal Regulations.

§ 812.110 Specific responsibilities of investigators.

(a) Awaiting approval. An investigator may determine whether potential subjects would be interested in participating in an investigation, but shall not request the written informed consent of any subject to participate, and shall not allow any subject to participate before obtaining IRB and FDA approval.

(b) Compliance. An investigator shall conduct an investigation in accordance with the signed agreement with the sponsor, the investigational plan, this part and other applicable FDA regulations, and any conditions of approval imposed by an IRB or FDA.

(c) Supervising device use. An investigator shall permit an investigational device to be used only with subjects under the investigator's supervision. An investigator shall not supply an investigational device to any person not authorized under this part to receive it.

(d) Financial disclosure. A clinical investigator shall disclose to the sponsor sufficient accurate financial information to allow the applicant to submit complete and accurate certification or disclosure statements required under part 54 of this chapter. The investigator shall promptly update this information if any relevant changes occur during the course of the investigation and for 1 year following completion of the study.

(e) Disposing of device. Upon completion or termination of a clinical investigation or the investigator's part of an investigation, or at the sponsor's request, an investigator shall return to the sponsor any remaining supply of the device or otherwise dispose of the device as the sponsor directs.

21 CFR Part 812: Investigational Device Exemptions, Subpart ERead on eCFR

Citations by year

Form 483 citations of 21 CFR 812.110 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 57 in 2010

Form 483 citations of 21 CFR 812.110 by inspection year
YearForm 483 citations
20087
200934
201057
201134
201237
201327
201442
201521
201614
201716
201820
201911
20203
20217
20225
20238
20248
20257
20267

How FDA words this citation

FDA's short descriptions and citation text on these 365 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Investigator non-compliance with agreement/plan/regulations315 citations
  • Inadequate financial disclosure by investigator27 citations
  • Subject participation prior to study approval16 citations
  • Investigator did not supervise use of investigational device4 citations
  • Informed consent obtained prior to study approval3 citations

Citation text

  • “An investigation was not conducted in accordance with the investigational plan.”112 citations
  • “An investigation was not conducted in accordance with the signed agreement and the investigational plan.”67 citations
  • “An investigation was not conducted in accordance with the investigational plan and conditions of approval imposed by an IRB.”27 citations
  • “An investigation was not conducted in accordance with the signed agreement, the investigational plan and applicable FDA regulations.”27 citations
Show 2 more
  • “The sponsor was not supplied sufficient financial information to allow submission of complete and accurate certification or disclosure statements.”16 citations
  • “An investigation was not conducted in accordance with conditions of approval imposed by an IRB.”15 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 812.110, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • Roderick H. Tung, M.D.2026-09-14 · FEI 3045746110Investigator non-compliance with agreement/plan/regulations
  • Jeffrey J. Sperinde, PhD2026-06-30 · FEI 3011936484Inadequate financial disclosure by investigator
  • Peter R. Movilla, MD2026-03-19 · FEI 3042665385Investigator non-compliance with agreement/plan/regulations
  • Stephane Leung Wai Sang, MD2026-02-24 · FEI 3041668878Investigator non-compliance with agreement/plan/regulations
  • Kevin E. Miller, MD2026-02-13 · FEI 3042247790Subject participation prior to study approval
  • Ryan D. Snowden, MD2026-01-16 · FEI 3015615983Investigator non-compliance with agreement/plan/regulations
Show 4 more
  • Ripul R. Panchal, DO2026-01-09 · FEI 3041295727Investigator non-compliance with agreement/plan/regulations
  • Alok Gambhir, MD, PhD.2025-10-01 · FEI 3039201029Investigator non-compliance with agreement/plan/regulations
  • Vincent DiGiovanni, DO2025-08-18 · FEI 3038623179Investigator non-compliance with agreement/plan/regulations
  • Hatice (NMI) Hasturk, DDS, Ph.D.2025-08-01 · FEI 3012169200Investigator non-compliance with agreement/plan/regulations

Often cited at the same inspection

Of the 348 inspections where FDA cited 21 CFR 812.110, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 812.110

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

Topics that count this section:Clinical investigators

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 812.110; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.