FDA has 5 inspections on file for Charles River Laboratories Cell Solutions Inc in Memphis, Tennessee (FEI 3007688680), from 2009 to 2017. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
5, 2009–2017
0 OAI · 1 VAI · 4 NAI
Inspections with 483 observations shown
1 of 5
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  2. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection of blood and blood components for further manufacturing purposes are not always followed.

    2. Donor deferral

      The standard operating procedure fails to include a written description of the donor deferral process.

    3. Expiration date

      The container label fails to include the expiration date, including the day, month, and year.

    4. Provide proper equipment to meet requirements

      Failure of equipment to perform in the manner for which it was designed so as to assure compliance with the official requirements prescribed in 21 CFR 606.

    5. Name, address, unique facility identifier

      The container label fails to include the name, address, and registration number.

    Show 5 more citations
    1. Donor notification

      The standard operating procedure fails to include a written description of the donor notification process and process for follow-up if the initial attempt at donor notification fails.

    2. Storage temperature in Celsius

      The container label fails to include the recommended storage temperature (in degrees Celsius).

    3. Donor suitability procedures not followed

      Failure to follow standard procedures and methods for determining the suitability of a donor as a source of blood.

    4. Infectious agent statement

      The container label fails to contain the statement: "This product may transmit infectious agents."

    5. Equipment observed, standardized, calibrated

      Equipment used in the collection of blood and blood components is not calibrated on a regularly scheduled basis as prescribed in the SOP Manual.

  4. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  5. NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone

Regulations FDA cited here

The 4 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 5 inspections.

Other FEIs recorded as Charles River Laboratories Cell Solutions Inc

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007688680 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007688680. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.