Apitoria Pharma Private Limited

Survey No. 10 & 13, Gaddapotharam Village, Ida-Kazipally, Jinnaram Mandal, Sangareddy District, India

FEI 3004086884DUNS 918917662

Company: Apitoria Pharma Private Limited · Linked by exact D-U-N-S match (861179030) in FDA registration data

FDA has 7 inspections on file for Apitoria Pharma Private Limited in Sangareddy District, India (FEI 3004086884), from 2011 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.

1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for API manufacturing and analysis, through Dec 31, 2027. Its registrant, Apitoria Pharma Private Limited (D-U-N-S 861179030), has registered 6 other sites.

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2011–2026
0 OAI · 6 VAI · 1 NAI
Inspections with 483 observations shown
3 of 7
Form 483s posted by FDA
↗
3 posted
Requested under FOIA
2 requests, latest Oct 2, 2024
Warning letters
None on file
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 10 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Equipment surface damage and residue

      Equipment used in the manufacturing of drug substances and in facilitate operations for its cleaning and maintenance. Specifically, during the day one walkthrough observed in the non-dedicated equipment used to manufacture drug substances for US market. A. Visible scratch marks and dents on the equipment surface of direct product contact were observed on the bottom section of equipment ID# (b)l, capacity (bT(4l, equipment ID# 003. Your maintenance check lists don't record the checking for scratch marks and dents. The latest preventive maintenance for (equipment ID# (b)(4) and (equipment ID# (b)l) was performed on 02 September 2024 and 26 August 2024, respectively. B. We observed a white residue buildup and what appeared to be scratch marks on the inside wall near the left hatch of "ready to use/ clean" equipment ID# CbT 00 (Capacity: I CbT'i) last cleaned on 12 September 2024 and equipment ID# 009, (capacity) last cleaned on 22 September 2024. We also observed white stains on the inside wall near the left hatch of "ready to use/ clean" equipment ID# 015 (Capacity Cb>1) last cleaned on 17 September 2024. (b)(4) (bT(4l (bT(4l (bT(4l

    2. Observation 2

      Incomplete manufacturing/cleaning instructions

      Failure to establish adequate manufacturing instructions in the production records for intermediates. Specifically, A. During two walkthroughs on 24 September 2024, at the (bY{] API stage for BPCR No. (b)(4) the batch record it states to 'an operator visually checks the (b)(4). However, such (b)(C4l was not included in the records. B. Your batch record for (bY{-4) for which drug substance manufacturing occurred at equipment (ID# (b)(4) did not include the operator instructions for the l--- equipment (ID# (b)(4). C. Your cleaning records for product change-over cleaning (ECR# ECR/381/PCO/007) from (b)(4) to (b)(4) drug substance (Batch# [illegible]) did not include the instructions, documented visual inspection checks on operation No. (b)(4). Additional cleaning changeover (ECR# ECR/381/PCO/003) and Batch# Cb><j did not include the instructions, documentation, and visual inspection checks on operation No. [illegible].

    Show observations 3 to 10
    1. Observation 3

      Inadequate training for personnel

      Failure to ensure training is regularly conducted by qualified individuals and that training covers particular operations that each employee performs. Specifically, A. Your firm fails to train your manufacturing personnel on cleaning verification of difficult to clean area of I (b)(4) equipment ID# CbT< 016 for cleaning verification after product-to-product change over cleaning (Type II cleaning) and periodic batch to batch cleaning (Type- III cleaning). B. Your firm fails to train your QC analysts on how to assess microbial growth in a growth promotion test using Cb 1 media.

    2. Observation 4

      Quality unit document review failures

      Failure of your quality unit to review and approve all appropriate quality-related documents. Specifically, I CbY{"l API (Lot# I CbTci has a change control (ID # CRF-UII-000693) due the size reduction of this batch from reduced batch size I CbY{j approved on 16 July 2024. BPCR review check list for this batch was reviewed and verified by quality assurance on 27 July 2024. However, the BPCR check list stating that any changed controls document was not initiated (Lot# I Cb)l1 and have attached to the investigation was initiated for not adequately reviewing quality documents.

    3. Observation 5

      Inadequate water system and sampling

      The water system design and the water sampling are not adequate. Specifically, A. The (b)l tank used in preparation of (b)(4) solution is located in an unprotected open environment without a sealed lid or a cover to protect the prepared solution from rain and other contaminants in the open air. B. During the water sampling demonstration, your operator did not demonstrate proper sampling technique. Specifically: 1. The operator failed to maintain disinfected clean state of the ends of the hose, and at one point dropped the hose such that the open end touched the floor. Later, the hose connectors were disinfected before sampling. 2. Your analyst failed to collect the (b)l4l water sample immediately after the removal of the protective covers from the (b)(4) sample port. 3. While sampling at point: (b)(4) the water was leaking from the (bT') connector due to a missing seal in the connector.

    4. Observation 6

      Inadequate change control and deviations

      Failure to establish and follow written procedures for intermediates and API batches to meet specifications. A. Your firm's change control request is inadequate and does not include risk evaluation and risk mitigation strategy. I CbTC deviation under Change Control Request CRF-UII-000286 (No. DE-UII-00001) cites size of approximately (Cb)l4l due to terminal process parameters and general process of the batch. The use of a reduced batch size was utilized that resulted in [illegible], despite the fact that the manufacturing process for batch size (bY{-4) was more suitable for the subject batch. B. Deviation DE-U08-001180 was initiated from R&D to Batch No. I CbT'i when unspecified impurities appeared in the chromatogram above the limit of NMT (Cb)C4l% for the total unknown impurities. The deviation was assigned as the root cause without confirmation at the proposed scale up batch size of (bY{-4).

    5. Observation 7

      Unqualified laminar flow hood

      Your equipment are not adequately qualified. Specifically, A. The Laminar Flow Hood (Property ID QLAM003) located in Microbiology lab is primarily used for preparation of (bY{-4) water samples for microbial count. Recently in 2024, the smoke studies conducted do not show laminar flow but instead the smoke streams are observed with downward trajectory from back end to front of the cabinet. B. The operating, cleaning and re-validation procedure does not include an equilibration period after initial startup and disinfection of the Laminar Flow Hood (Property ID QLAM003).

    6. Observation 8

      Inadequate cleaning validation protocol

      Your equipment cleaning validation protocol is inadequate. Specifically, Your firm uses non-dedicated equipment for manufacturing US and non-US products. During the product to-product changeover equipment cleaning (Type II cleaning) validation for API in Block (ID module) (b)(4) and US product discrepancies were noted. Non-US product (bY{1) API in Block (b)(4) cleaning validation was completed on 18 May (protocol # [illegible]). For Swab limit (ppm), you have calculated total surface area (TSA) of all the equipment surface area and used (bY{-4)% for additional surface area (area of pipes, tubes, spatulas, etc.). During the equipment cleaning validation completed on 19 June 2024 (protocol# P-CLV-01-00), for Swab limit (ppm) you have calculated TSA by addition of all the equipment surface area and then reduced the additional surface area to (b)(4)% which ultimately reduced the TSA. Your calculation methodology is not supported with justification, change controls or SOP (Equipment cleaning validation, CQA045) to support the above change in equipment cleaning validation protocol.

    7. Observation 9

      Unmapped/undefined equipment qualification

      Failure to ensure that new and modified facilities are qualified. A. During the review of installation qualification of (bT'i) (Set point 2°C to 8°C; per (b)(4)) your firm has not conducted the temperature mapping for worst case (i.e. fully loaded) condition. B. SOP CQA020, titled "Calibration of humidity, cooling incubators and deep freezers", updated October 2023, stated that mapping of cooling incubators should be done but the worst-case conditions have not been specified in the SOP.

    8. Observation 10

      Uncontrolled label issuance

      Failure of your quality unit to establish a system of control for labeling and packaging materials. Specifically, Your firm has generated labels for "retest due" and these labels were issued outside from the established ERP (Enterprise resource plan). No issuance logs and label reconciliation process was in place for these labels.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 2 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Inadequate cleaning procedures/validation

      Written procedures are not established to assure that equipment is clean prior nducted between production batches. Specifically, the cleaning validation is not adequate in that: ) the cleaning procedures currently in use, Equipment Cleaning /04/14, were not evaluated in the cleaning validation for the as no document control number. ) the cleaning validation report CV/NOI.AXJl-02, dated 4/02/14, (b)(4) (Equipment ID (b)(4) 013). ) the cleaning validation report CV/NOI.AXJl-02, dated 4/02/14, Equipment Cleaning and Use Log and the logbook was reportedly not maintained, therefore the details pertaining to the actual cleaning of the equipment could not be verified..

    2. Observation 2

      Incomplete investigations of deviations

      Investigations into critical deviations are not always thorough including appropriate Specifically, investigations pertaining to out--of-specification (b)(4) API, which were performed for several batches, were not thorough. For example, ) Investigation into OOS-UOS-000734, OOS-UOS-000735, OOT-U08-000151, reporting OOS for Residual Solvents (b)(4), concluded that however; there is no scientific data to support this conclusion and all potential root causes were not investigated or discussed. Additionally, the investigation did not include a review (b)(4) 002), which was performed by the previous site owner quality Unit for its acceptability. Additionally, the firm does not ) Investigation into OOS-UOS-000860 & OOS-U08-000861 initiated included that a malfunction pertaining to leakage in the (b)(4) were not investigated and/or discussed. ) Deviation Record Number DE~UOS-000455 initiated 12/19/ reprocessed Batch (b)(4) (reprocessed batch of (b)(4)) condensation in the packing material during storage of the input were not investigated and/or discussed. The responsibilities and procedures applicable to the Quality Specifically, procedures are not established to determine when a reprocessed batch meets its quality attributes throughout the retest or expiry period, when no changes will (b)(4) API-Reprocessed, including (b)(4) Batch (batch (b)(4)) was not put on stability, and there is no Batch production and control records do not always include control of each batch. For example, electronic batch processing data for manufacture of (b)(4), including (b)(4) ...

  4. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  5. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 1 observation

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Inadequate investigations

      lnvestigations afe inadequate, Specially, daviealiaA— DR ~Uos-coeR was ifaidiared on 22-FUAe-201S becaise none of +he Holloring (AP) Batches were Suen bed For Shalili by I BR "oon oad Conkmin> Ad. "Numbar of Pleviccs daviedisms Tm Haie Peadock 2 Syste". This feld cequizes provievs devisbions of a Sema Prrduct of dauiekion type +e be reporied, ne previous Saviariead were feporbed Tn Huis Lield. Oeniarion - DR -UOE-cocer was reporied aPP@xTmmiels 2 montis previsusly For failiag to sie vrass or EEE THEY js replat DNEwansies Sheth ar oy Fnspeciions RS— BV YE PT ER ties Fm TOA 83 (WOR) Prous ROT GROUT INSPECTIONAL OBSERVATIONS mae vor |

  6. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  7. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Oct 2, 2024. Each one asked for a Form 483.

  • Inspection ending Sep 27, 2024VAI

    1 request. In the inspection history above. FDA has posted its Form 483.

    1. Oct 1, 2024FDA-FOIA-2024-8560Form 483Closed
  • Inspection date not given

    1 request. None of FDA’s log entries for these requests say records were released.

    1. Oct 2, 2024FDA-FOIA-2024-8603Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3004086884, one row per refused product line.

1 refusal on 2017-05-26 · charged DIRECTIONS on 1 · no FDA sample analysis

FDA firm profile
  • 2017-05-26

    EXHIBITS - OTHER DRUG RELATED ITEMS N.E.C.66YIY99

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    Division of Northeast Imports201-3007858-7/11/1

FDA drug establishment registration

FDA registration search
Registered operations
API manufacturing, analysis
Registration runs through
Dec 31, 2027
Registrant
Apitoria Pharma Private LimitedD-U-N-S 861179030

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Other sites with the same registrant

FDA’s registration file names Apitoria Pharma Private Limited (D-U-N-S 861179030) as the registrant for this site and the 6 others below. That exact D-U-N-S number is the only link. A registrant is the firm that filed the registration, which is not always the owner.

Other FEIs recorded as Apitoria Pharma Private Limited

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. This site’s records are filed under FEI 3004086884 in FDA’s data; other sites under its registrant are listed in summary only. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every inspection and enforcement record on this page is filed under FEI 3004086884. Other sites filed under the same registrant D-U-N-S are listed in summary only. Records filed under another FEI or name are otherwise not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.