FDA has 2 inspections on file for Allstar Marketing Group Llc in Hawthorne, New York (FEI 3003799587), from 2022 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA posted the Form 483 from that inspection. FDA published 6 citations from this inspection. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
2, 2022–2026
0 OAI · 1 VAI · 1 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
↗
Requested under FOIA
2 requests, latest May 13, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesPosted 483 · 6 citations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.1

      Records of risk management activities in product realization not maintained

      Records of risk management activities in product realization were not maintained.

    2. ISO 13485:2016 Clause 7.4.3

      Purchased product verification activities are inadequate

      The extent of verification activities for purchased product was not based on the supplier evaluation results and proportionate to the risks associated with the purchased product.

    3. Records for complaints involving failure to meet specifications not maintained

      Records of the review, evaluation, and investigation for any complaints involving the possible failure of a device, labeling, or packaging to meet any of its specifications were not adequately maintained.

    4. ISO 13485:2016 Clause 4.2.1

      QMS documentation does not include required documents

      The quality management system documentation does not include required procedures and records.

    5. ISO 13485:2016 Clause 7.4.1

      Criteria for the evaluation and selection of suppliers not established

      Criteria for the evaluation and selection of suppliers were not adequately established.

    Show 1 more citation
    1. Report of death, injury

      The importer failed to submit a report to FDA, with a copy to the manufacturer, within 30 days after receiving information that one of its marketed devices may have caused or contributed to a serious injury.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on May 13, 2026. Each one asked for a Form 483.

  • Inspection ending Mar 11, 2026VAI

    1 request. In the inspection history above. FDA has posted its Form 483.

    1. May 13, 2026FDA-FOIA-2026-4917Form 483· 1 of 5 sitesRecords released
  • Inspection ending Mar 21, 2022NAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Apr 27, 2022FDA-FOIA-2022-3030Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 3 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
  • 21 CFR 803.40If I am an importer, what reporting requirements apply to me?1 of 2
  • 21 CFR 820.35Control of records1 of 2

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003799587 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003799587. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.