FDA has 4 inspections on file for Control Micro Systems, Inc. in Winter Park, Florida (FEI 3002921502), from 2009 to 2018. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2009–2018
- 0 OAI · 1 VAI · 3 NAI
- Inspections with 483 observations shown
- 1 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Aperture label - laser radiation
The warning statement "AVOID EXPOSURE---Laser radiation is emitted from this aperture" was not affixed in close proximity to each aperture through which is emitted accessible laser or collateral radiation in excess of the accessible emission limits of Class I and table VI of Section 1040.10(d).
Basis for selecting methods, devices, procedures
You have not established and maintained records of the basis for selecting the methods, devices and procedures used in tests for electronic product radiation safety.
Class IV "Danger" label
Each Class IV laser product does not have affixed a label bearing the DANGER logotype specified in the regulation, bearing the wording "LASER RADIATION--AVOID EYE OR SKIN EXPOSURE TO DIRECT OR SCATTERED RADIATION" in the position specified in the regulation and bearing the wording "CLASS IV LASER PRODUCT" in the position specified in the regulation.
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
Regulations FDA cited here
The 2 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 4 inspections.
- 21 CFR 1002.30Records to be maintained by manufacturers1 of 1
- 21 CFR 1040.10Laser products1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002921502 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002921502. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.