Skip to content

Titan USA, LLC formerly Amstech

13444 Hipworth Rd, Milton, Georgia, United States

FEI 3002828006

FDA has 2 inspections on file for Titan USA, LLC formerly Amstech in Milton, Georgia (FEI 3002828006), from 2009 to 2012. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2009–2012
1 OAI · 1 VAI · 0 NAI
Inspections with 483 observations shown
1 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Radiation Control and Health Safety ActCitations not published
  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Remote interlock connector

      Each laser system classified as a Class IIIb or Class IV laser product was not incorporated with a readily available remote interlock connector having an electrical potential difference of no greater than 130 root-mean-square volts between terminals.

    2. Key control

      Each laser system classified as a Class IIIb or Class IV] laser product did not have a removable key

    3. Beam attenuator

      Each laser system classified as a Class II, III, or IV laser product, was not provided with one or more permanently attached means, other than laser energy source switch(es), electrical supply main connectors, or the key-actuated master control, capable of preventing access by any part of the human body to all laser and collateral radiation in excess of the accessible emission limits of Class I and table VI.

    4. User info - list controls, etc.

      The manufacturer did not provide or cause to be provided a listing of the warning: "Caution --- use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.".

    5. User information, adequate instructions

      Adequate instructions for each laser product were not provided or caused to be provided for clear warnings concerning precautions to avoid possible exposure to laser and collateral radiation in excess of the accessible emission limits specified in the regulations.

    Show 1 more citation
    1. Methods, devices, procedures for testing

      You have not established and maintained records of the devices and procedures used in tests for electronic product radiation safety.

Regulations FDA cited here

The 2 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002828006 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002828006. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.