GAM Laser, Inc.
7100 Tpc Dr Ste 200, Orlando, Florida, United States
FEI 3001705846
FDA has 3 inspections on file for GAM Laser, Inc. in Orlando, Florida (FEI 3001705846), from 2010 to 2018. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2010–2018
- 0 OAI · 1 VAI · 2 NAI
- Inspections with 483 observations shown
- 1 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Methods, devices, procedures for testing
You have not established and maintained records of the methods, devices and procedures used in tests for electronic product radiation safety.
User information, radiation output information
A statement of the magnitude of the maximum radiant power and maximum radiant energy per pulse of the accessible laser radiation detectable in each direction in excess of the accessible emission limits in the standard was not provided or caused to be provided for each laser product.
User info - list controls, etc.
The manufacturer did not provide or cause to be provided a listing of the warning: "Caution --- use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.".
Reproduction of affixed information
The manufacturer did not provide or cause to be provided in all catalogs, specification sheets and descriptive brochures pertaining to each laser product, a legible reproduction of the class designation and the warning required b to be affixed to the product.
User information, adequate instructions
Adequate instructions for each laser product were not provided or caused to be provided for operation and clear warnings concerning precautions to avoid possible exposure to laser and collateral radiation in excess of the accessible emission limits specified in the regulations.
Show 1 more citation
Failure to submit
You did not submit an annual report for products requiring one.
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
Regulations FDA cited here
The 3 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 3 inspections.
- 21 CFR 1002.13Annual reports1 of 1
- 21 CFR 1002.30Records to be maintained by manufacturers1 of 1
- 21 CFR 1040.10Laser products1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001705846 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001705846. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.