FDA has 3 inspections on file for GAM Laser, Inc. in Orlando, Florida (FEI 3001705846), from 2010 to 2018. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 6 citations from this inspection. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2010–2018
0 OAI · 1 VAI · 2 NAI
Inspections with 483 observations shown
1 of 3
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Methods, devices, procedures for testing

      You have not established and maintained records of the methods, devices and procedures used in tests for electronic product radiation safety.

    2. User information, radiation output information

      A statement of the magnitude of the maximum radiant power and maximum radiant energy per pulse of the accessible laser radiation detectable in each direction in excess of the accessible emission limits in the standard was not provided or caused to be provided for each laser product.

    3. User info - list controls, etc.

      The manufacturer did not provide or cause to be provided a listing of the warning: "Caution --- use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.".

    4. Reproduction of affixed information

      The manufacturer did not provide or cause to be provided in all catalogs, specification sheets and descriptive brochures pertaining to each laser product, a legible reproduction of the class designation and the warning required b to be affixed to the product.

    5. User information, adequate instructions

      Adequate instructions for each laser product were not provided or caused to be provided for operation and clear warnings concerning precautions to avoid possible exposure to laser and collateral radiation in excess of the accessible emission limits specified in the regulations.

    Show 1 more citation
    1. Failure to submit

      You did not submit an annual report for products requiring one.

  2. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
  3. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone

Regulations FDA cited here

The 3 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 3 inspections.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3001705846 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3001705846. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.