Zeus Scientific, Inc.
199 Evans Way, Branchburg, New Jersey, United States
FEI 2242436
FDA has 4 inspections on file for Zeus Scientific, Inc. in Branchburg, New Jersey (FEI 2242436), from 2009 to 2020. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2009–2020
- 0 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch Monitoring2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Not exempt--lacks For Investigational Use statement
A delivery for product testing of an in vitro diagnostic product prior to full commercial marketing, is not exempt from the requirements for labels and accompanying labeling because all labeling does not bear a prominently placed statement "For Investigational Use Only. The performance characteristics of this product have not been established.".
IRB review requirement
A clinical investigation requiring prior submission to the FDA was initiated without IRB approval .
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training
Employees have not been adequately trained.
Management review
The procedures addressing submission of information on identified quality problems and corrective and preventive actions for management review were not complete.
Statistical methodology
Corrective and preventive action procedures addressing the use of appropriate statistical methodology to identify existing and potential causes of nonconforming product or other quality problems were not complete.
Analysis of data sources
Appropriate sources of quality data are not adequately analyzed to identify existing and potential causes of nonconforming product and other quality problems.
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.25Personnel1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 56.103Circumstances in which IRB review is required1 of 2
- 21 CFR 809.10Labeling for in vitro diagnostic products1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2242436 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2242436. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.