Skip to content

DLP Twin County Regional Healthcare, LLC

200 Hospital Dr, Galax, Virginia, United States

FEI 1176729

FDA has 3 inspections on file for DLP Twin County Regional Healthcare, LLC in Galax, Virginia (FEI 1176729), from 2009 to 2013. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 10 citations from this inspection. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2009–2013
0 OAI · 1 VAI · 2 NAI
Inspections with 483 observations shown
1 of 3
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Thorough investigations

      Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy and a failure of a lot or unit to meet any of its specifications.

    2. Equipment observed, standardized, calibrated

      Equipment used in the compatibility testing of blood and blood components is not calibrated on a regularly scheduled basis as prescribed in the SOP Manual.

    3. Biological product deviation report

      Failure to submit a biological product deviation report.

    4. Required records - transfusion reaction and complaints

      Failure to maintain records of transfusion reaction reports and complaints, including investigation and follow up.

    5. Record review prior to release

      All records pertinent to a lot or unit were not reviewed before the release or distribution of a lot or unit of final product.

    Show 5 more citations
    1. Equipment calibration frequency

      Equipment used in the collection of blood and blood components is not observed, standardized and calibrated with at least the frequency required.

    2. Donor suitability not performed on day of collection

      Failure to determine the suitability of a donor as a source of blood by a qualified physician or by a trained person under his supervision on the day of collection.

    3. Caution statement - transfusion

      The product is intended for transfusion and the container label fails to include the statement(s) ?Rx only?.

    4. Legibility and indelibility

      Records are not indelible.

    5. Qualifications of responsible personnel

      The personnel responsible for the collection, processing, compatibility testing, storage, and distribution of blood or blood components are not adequate in training and experience, including professional training as necessary to assure competent performance of their assigned functions, and to ensure that the final product has the safety, purity, potency, identity and effectiveness it purports or is represented to possess.

  2. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  3. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 3 inspections.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1176729 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1176729. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.