Walter L. Shepeard Community Blood Center, Inc.
1533 Wrightsboro Rd, Augusta, Georgia, United States
FEI 1046616
FDA has 8 inspections on file for Walter L. Shepeard Community Blood Center, Inc. in Augusta, Georgia (FEI 1046616), from 2009 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 606.100(b) and 21 CFR 606.100(c).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 8, 2009–2025
- 0 OAI · 4 VAI · 4 NAI
- Inspections with 483 observations shown
- 3 of 8
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- NAINo objectionable conditions that need follow-upBiologics · Vaccines and Allergenic ProductsNone
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Written methods for investigating adverse reactions
The standard operating procedure fails to include a written description of the procedures for investigating adverse donor and recipient reactions.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Provisions to monitor lab test procedures & instruments
Failure to establish adequate provisions for monitoring the reliability, accuracy, and performance of laboratory test procedures and instruments.
Documentation
Failure to document that you have successfully notified a deferred donor.
Physician notification
Failure to provide to an autologous donor's referring physician information that the autologous donor is deferred based on the results of tests for evidence of infection due to communicable disease agents, and the reason for that decision.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the processing of blood and blood components for homologous transfusion, autologous transfusion, and further manufacturing purposes are not always maintained and followed.
Personnel capabilities
Failure to assure that personnel have the necessary training in and a thorough understanding of the operations which they perform.
Show 2 more citations
Thorough investigations
Failure to perform a thorough investigation of an unexplained discrepancy.
Following manufacturer's instructions
Failure to use supplies and reagents in a manner consistent with instructions provided by the manufacturer.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Donor suitability procedures not followed
Failure to follow standard procedures and methods for determining the suitability of a donor as a source of blood.
Qualifications of responsible personnel
The personnel responsible for the collection of blood or blood components are not adequate in training and experience, including professional training as necessary to assure competent performance of their assigned functions, and to ensure that the final product has the safety, purity, potency, identity and effectiveness it purports or is represented to possess.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection, processing, and distribution of blood and blood components for homologous transfusion, autologous transfusion, and further manufacturing purposes are not always followed.
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.
Concurrent documentation
Records are not concurrently maintained with the performance of each significant step in the collection of each unit of blood and blood components so that all steps can be clearly traced.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 606.100Standard operating procedures3 of 3
- 21 CFR 600.10Personnel1 of 3
- 21 CFR 606.20Personnel1 of 3
- 21 CFR 606.65Supplies and reagents1 of 3
- 21 CFR 606.140Laboratory controls1 of 3
- 21 CFR 606.160Records1 of 3
- 21 CFR 630.40Requirements for notifying deferred donors1 of 3
- 21 CFR 640.31 of 3
Other FEIs recorded as Walter L. Shepeard Community Blood Center, Inc.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Walter L. Shepeard Community Blood Center, Inc.FEI 3005209270Evans, Georgia, United States
Last inspected 2026-05-13NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1046616 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1046616. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.