Hoxworth Blood Center University of Cincinnati
2220 Grandview Dr Ste 140, Fort Mitchell, Kentucky, United States
FEI 1000149727
FDA has 8 inspections on file for Hoxworth Blood Center University of Cincinnati in Fort Mitchell, Kentucky (FEI 1000149727), from 2009 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 8, 2009–2026
- 0 OAI · 1 VAI · 7 NAI
- Inspections with 483 observations shown
- 1 of 8
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Determination of lot numbers and supplies
Appropriate records are not available to determine the lot numbers of supplies used for specific units of the final product.
Arm preparation
The phlebotomy site is not prepared by a method that gives maximum assurance of a sterile container of Whole Blood.
Provide space for examination
Failure to provide adequate space for private examinations of individuals to determine their suitability as blood donors.
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
- NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
Regulations FDA cited here
The 3 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 8 inspections.
- 21 CFR 606.40Facilities1 of 1
- 21 CFR 606.160Records1 of 1
- 21 CFR 640.4Collection of the blood1 of 1
Other FEIs recorded as Hoxworth Blood Center University of Cincinnati
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Hoxworth Blood Center University of CincinnatiFEI 1570969Cincinnati, Ohio, United States
Last inspected 2025-09-08NAI
Hoxworth Blood Center University of CincinnatiFEI 3005890890West Chester, Ohio, United States
Last inspected 2024-02-28NAI
Hoxworth Blood Center University of CincinnatiFEI 1527551Cincinnati, Ohio, United States
Last inspected 2016-10-06VAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000149727 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000149727. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.