Skip to content

2026-W33 · 2026-08-10 to 2026-08-16

FDA enforcement, week of 2026-08-10

In 2026-W33 (2026-08-10 to 2026-08-16), FDA posted 2 warning letters from its drug, biologics and device offices, and listed 1 Form 483 in its Electronic Reading Room for the first time. FDA's Enforcement Report added 84 drug and device recalls, 6 of them Class I. FDAgent recorded 1 new Official Action Indicated (OAI) classification in FDA's inspection data. FDA published 8 firm entries on its import alert lists, 6 of them on a Red List.

Warning letters posted
2
Posted 2026-08-11
Form 483s listed
1
In the reading room
Recalls
84
6 Class I · 75 II · 3 III
New OAI classifications
1
Newly listed by FDA

Ask FDAgent about this week

FDAgent reads the warning letters FDA posted this week and answers in plain English, with links to the letters.

Edit it or ask as written. 3 questions free, no card needed.

Warning letters

2 letters FDA posted on its warning letters page this week, from its drug, biologics and device offices. Each letter's issue date can be weeks before its posting date. Warning letters last synced Nov 23, 2026, 04:17 UTC.

An FEI links to its facility page only when the letter names exactly one FEI and FDAgent holds FDA records for it.

Import alert entries

8 firm entries FDA published on its import alert pages this week, dated by each entry's date published: 6 on a Red List, 2 on a Green List. FDA lists firms by name and address; FDAgent links an entry to a facility only when FDA printed the firm's FEI. 1 entry whose name may be a person's rather than a firm's is not listed or counted here; signed-in users can see them in FDAgent. Import alerts last synced Nov 23, 2026, 04:17 UTC.

66-40Red List3 entriesShow allDetention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPsAlert on fda.gov
  • Published 2026-08-14

    TSP Solutions Pte. Ltd.

    SINGAPORE

  • Published 2026-08-13

    Rohto-Mentholatum (Vietnam) Co., Ltd.

    VIETNAM

  • Published 2026-08-11

    Shoolin Pharma Chem LLP

    INDIA

66-41Red List1 entryShow allDetention Without Physical Examination of Unapproved New Drugs Promoted In The U.S.Alert on fda.gov
  • Published 2026-08-13

    Global Vision, Inc.

    JAPAN

76-01Green List2 entriesShow allDetention Without Physical Examination Of Medical Instruments from PakistanAlert on fda.gov
  • Published 2026-08-13

    Sobytek (PVT) Ltd

    PAKISTAN

  • Published 2026-08-13

    Surgicell Industries

    PAKISTAN

95-05Red List2 entriesShow allDetention Without Physical Examination of Electronic Products that Fail to Comply with Performance Standards or to Have Required CertificationAlert on fda.gov
  • Published 2026-08-10

    Shenzhen Mapeta Electronic Co. Ltd

    CHINA

  • Published 2026-08-10

    SHENZHEN SCULPFUN TECHNOLOGY CO LTD

    CHINA

Form 483s listed in FDA's Electronic Reading Room

1 Form 483 FDA listed for the first time in its Electronic Reading Room this week, dated by the publish date FDA gives each. FDA sometimes lists a 483 days after that date, so a week can still gain 483s after it ends. FDA posts only some 483s, often months after the inspection. Reading room last synced Nov 23, 2026, 04:33 UTC.

Recalls by class

Drug and device recalls in FDA's Enforcement Report dated this week, from openFDA, which FDA last updated Nov 11, 2026.

Class I6 recallsShow all
  • Z-2796-2026

    AVID Medical, Inc.

    Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05

    Reason: Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and…

    Device · initiated 2026-04-30 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2797-2026

    AVID Medical, Inc.

    Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. SAMMC ANGIOGRAPHY SAMM066-15

    Reason: Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and…

    Device · initiated 2026-04-30 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2798-2026

    AVID Medical, Inc.

    Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. HEART CATH, SELF REGIONAL SELF131-05

    Reason: Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and…

    Device · initiated 2026-04-30 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2879-2026

    Boston Scientific Neuromodulation Corporation

    INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.

    Reason: Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.

    Device · initiated 2026-06-17 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • D-0743-2026

    Sunny Pharmtech Inc.

    Cyclophosphamide for Injection, USP, 1 gram/vial, One Single-Dose Vial, Rx only, Manufactured for: Long Grove Pharmaceuticals, LLC, Rosemont, IL 60018. Manufactured in Taiwan, NDC 81298-8112-1

    Reason: Presence of Particulate Matter; identified as stainless steel

    Drug · initiated 2026-07-17 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • D-0744-2026

    Sunny Pharmtech Inc.

    Cyclophosphamide for Injection, USP, 2 gram/vial, One Single-Dose Vial, Rx only, Manufactured for: Long Grove Pharmaceuticals, LLC, Rosemont, IL 60018. Manufactured in Taiwan, NDC 81298-8114-1

    Reason: Presence of Particulate Matter; identified as stainless steel

    Drug · initiated 2026-07-17 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
Class II75 recallsShow all
  • Z-2894-2026

    Aniara Diagnostica LLC

    5-PLASMA Bivalirudin AMR Verification Kit, Model Number 5D-45905

    Reason: Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values.

    Device · initiated 2026-07-01 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2895-2026

    Aniara Diagnostica LLC

    5-PLASMA Bivalirudin Calibrator, Model Number 5D-46905

    Reason: Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values.

    Device · initiated 2026-07-01 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2896-2026

    Aniara Diagnostica LLC

    5-PLASMA Bivalirudin Control, Model Number 5D-44905

    Reason: Stability issues with the reconstituted lyophilized human plasma and bivalirudin matrix have been identified. These stability issues lead to a degradation of measurable bivalirudin in the controls, calibrators, and linearity values.

    Device · initiated 2026-07-01 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • D-0733-2026

    Aurobindo Pharma USA Inc

    Dicyclomine Hydrochloride Capsules, USP, 10mg, 1,000 bottles, Rx only, Distributed by: Aurobindo Pharma USA, Inc., 279 Princeton-Hightstown Road, East Windsor, NJ 08520, Made in India, NDC 59651-719-99

    Reason: shortfill; reports of empty capsules.

    Drug · initiated 2026-07-23 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2834-2026

    Beckman Coulter Biomedical GmbH

    REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections

    Reason: Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes.

    Device · initiated 2026-04-29 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2888-2026

    Beckman Coulter Inc.

    REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 Micr…

    Reason: Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."

    Device · initiated 2026-06-26 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2889-2026

    Beckman Coulter Inc.

    REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway…

    Reason: Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."

    Device · initiated 2026-06-26 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2877-2026

    Biofire Defense

    FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system. The FilmArray NGDS Warrior Panel detects and…

    Reason: Firm has identified mismatches between the outbreak strain and the reference strain used in the development of its qualitative, multiplexed, nucleic acid-based in vitro diagnostic panel test.

    Device · initiated 2026-07-16 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2867-2026

    Biomet, Inc.

    Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary Bone Saw Blade Assembly 12mm

    Reason: Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade.

    Device · initiated 2026-07-01 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2868-2026

    Biomet, Inc.

    Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475625 Product Description: Intramedullary Bone Saw Blade Assembly 13mm

    Reason: Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade.

    Device · initiated 2026-07-01 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2869-2026

    Biomet, Inc.

    Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475630 Product Description: Intramedullary Bone Saw Blade Assembly 14mm

    Reason: Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade.

    Device · initiated 2026-07-01 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2870-2026

    Biomet, Inc.

    Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475635 Product Description: Intramedullary Bone Saw Blade Assembly 15mm

    Reason: Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade.

    Device · initiated 2026-07-01 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2871-2026

    Biomet, Inc.

    Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary Bone Saw Blade Assembly 16mm

    Reason: Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade.

    Device · initiated 2026-07-01 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2838-2026

    Boston Scientific Corporation

    Radial Jaw 4 Standard Capacity Biopsy Forceps with the following descriptions and UPN: 1. Radial Jaw 4 SC 240cm Box 20; Outer Box UPN: M00513401; Inner Box UPN: M00513400. 2. Radial Jaw 4 SC 240cm Box 40; Outer Box UPN: M00513402; Inner Box UPN: M00513400. 3. Radial Jaw 4 SC w/…

    Reason: Potential for breach in sterile barrier of device packaging.

    Device · initiated 2026-06-30 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2839-2026

    Boston Scientific Corporation

    Radial Jaw 4 Large Capacity Biopsy Forceps with the following descriptions and UPN: 1. Radial Jaw 4 LC 240cm Box 20; Outer Box UPN: M00513321; Inner Box UPN: M00513320. 2. Radial Jaw 4 LC 240cm Box 40; Outer Box UPN: M00513322; Inner Box UPN: M00513320. 3. Radial Jaw 4 LC w/Need…

    Reason: Potential for breach in sterile barrier of device packaging.

    Device · initiated 2026-06-30 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2840-2026

    Boston Scientific Corporation

    Radial Jaw 4 Jumbo Biopsy Forceps with the following descriptions and UPN: Radial Jaw 4 Jumbo 240cm Box 5; Outer Box UPN: M00513363; Inner Box UPN: M00513360.

    Reason: Potential for breach in sterile barrier of device packaging.

    Device · initiated 2026-06-30 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2911-2026

    Contego Medical, Inc

    Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40

    Reason: Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be los…

    Device · initiated 2026-06-29 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2854-2026

    Copan WASP

    Brand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.33.1, 2.33.3 and 2.34.2 Product Description: WASPLab is an automated device intended to be used for the incubation and the digital imaging of agar culture plates streaked with…

    Reason: Due to a software issue within the WebApp that potentially prevents the user to overwrite the results.

    Device · initiated 2026-06-16 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2909-2026

    Freudenberg Medical, LLC

    Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LEF

    Reason: Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.

    Device · initiated 2026-06-17 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2910-2026

    Freudenberg Medical, LLC

    Blom-Singer Classic Indwelling Voice Prosthesis, 16Fr/10 mm REF: 1610-LF

    Reason: Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.

    Device · initiated 2026-06-17 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2876-2026

    GE Medical Systems Information Technologies Inc

    GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph

    Reason: Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrec…

    Device · initiated 2026-06-29 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2878-2026

    GE Medical Systems Information Technologies Inc

    GE Healthcare Portrait Core Services SW V 1.2.0, REF 5700592. patient monitoring system

    Reason: A software vulnerability may allow unauthorized persons to gain privileged access to the device, which could result in loss of patient monitoring, missed patient alarms, and/or access to Protected Health Information (PHI).

    Device · initiated 2026-07-10 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2882-2026

    Happiest Baby, Inc

    Happiest Baby, Inc Pre-loved SNOO Smart Sleeper Model:S1000

    Reason: This recall is retroactively being reported. Refurbished bassinets failed to meet quality standards prior to distribution.

    Device · initiated 2024-05-15 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2887-2026

    Hardy Diagnostics

    HardyDisk AST Cefiderocol 30¿g (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30¿g (FDC30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of Enterobacterales, Pseudomonas aeruginosa, Acinetobacter baumannii com…

    Reason: Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility results and may lead to selection of an ineffective antimicrobial therapy, potentially resulting in delayed appropriate treatment, prolong…

    Device · initiated 2026-06-08 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2859-2026

    Innovasis, Inc

    Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System

    Reason: Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod…

    Device · initiated 2026-05-22 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2899-2026

    Instrumentation Laboratory

    GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized who…

    Reason: The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormaliti…

    Device · initiated 2026-05-21 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2900-2026

    Instrumentation Laboratory

    GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 00055445010 00055445011 00055360010.A

    Reason: The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their s…

    Device · initiated 2026-05-21 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2836-2026

    INTEGRATED MEDCRAFT LLC

    FORWARD Oral Airway Kit (OPA Sizes 5, 6.5, 7, 8, 9, 10, 12). REF Number: 1011-40002. Containing the affected One-piece Guedel airway components: 1. Size 00, ISO 5.0, blue. Product Code: 1100050. 2. Size 000, ISO 3.5, light green. Product Code: 1000035.

    Reason: Devices could potentially contain an occluded or partially occluded Guedel Airway tip.

    Device · initiated 2026-07-07 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2837-2026

    INTEGRATED MEDCRAFT LLC

    Kingfisher Compact Individual First Aid Kit. REF Number: 76656. Containing the affected One-piece Guedel airway components: 1. Size 00, ISO 5.0, blue. Product Code: 1100050. 2. Size 000, ISO 3.5, light green. Product Code: 1000035.

    Reason: Devices could potentially contain an occluded or partially occluded Guedel Airway tip.

    Device · initiated 2026-07-07 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2860-2026

    Maquet Critical Care AB

    Flow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 6677300 Flow-i C40 6677400 Flow-i C20 6888520 Flow-i C30 6888530 Flow-i C40 6888540 Flow-c 6887700 Flow-e 6887900

    Reason: Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.

    Device · initiated 2026-06-24 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2858-2026

    Medical Action Industries, Inc. 306

    Brand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Number: 78791

    Reason: Convenience kits contain needle free access device that has an improper seal which may result in doming or tearing of the silicone seal.

    Device · initiated 2026-05-21 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2874-2026

    Merge Healthcare, Inc.

    Merge PACS, medical imaging diagnostic workstation

    Reason: Thumbnails from a previously opened patient study may be displayed in the Thumbnail Bar while a different patient study is currently open.

    Device · initiated 2026-06-30 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2875-2026

    Merge Healthcare, Inc.

    Merge OrthoPACS, medical imaging diagnostic workstation

    Reason: Thumbnails from a previously opened patient study may be displayed in the Thumbnail Bar while a different patient study is currently open.

    Device · initiated 2026-06-30 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2856-2026

    MUNEVO GMBH

    munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System

    Reason: A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in l…

    Device · initiated 2026-06-19 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2861-2026

    Nuclein LLC

    Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002 Software Version: 6.0.2 Product Description: The DASH¿ Rapid PCR Instrument (DASH Instrument) is an automated, fluorescence-based, reverse-transcription polymerase chain r…

    Reason: Due to a software issue which may result in false co-infection (e.g., Positive for Flu A, when SARS-CoV-2 is present).

    Device · initiated 2025-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2857-2026

    Oxos Medical Inc

    Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point o…

    Reason: Due to a defective circuit board that overheats

    Device · initiated 2026-06-11 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2891-2026

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

    Philips Allura under the following System Descriptions and corresponding model numbers: 1. Allura Xper FD10; Model Numbers: 722003, 722010, 722026. 2. Allura Xper FD10 OR Table; Model Number: 722033. 3. Allura Xper FD10/10; Model Numbers: 722005, 722011, 722027. 4. Allura Xper F…

    Reason: Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.

    Device · initiated 2026-06-17 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2892-2026

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

    Philips Azurion under the following System Descriptions and corresponding model numbers: 1. Azurion 3 M12; Model Numbers: 722063, 722221. 2. Azurion 3 M15; Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12; Model Number: 722227. 4. Azurion 5 M20; Model Numbers: 722228, 722…

    Reason: Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.

    Device · initiated 2026-06-17 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2893-2026

    PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

    Philips Allura, Cardio Vascular-Allura Centron. Model Number: 722400. Interventional Fluoroscopic X-Ray System.

    Reason: Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.

    Device · initiated 2026-06-17 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2841-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: Quantum 4Front 2 Model/Catalog Number: Quantum 4Front 2 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2842-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: Quantum Outback Model/Catalog Number: Quantum Outback Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2843-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Stretto Model/Catalog Number: Q6 Edge 3 Stretto Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2844-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: R-TRAK Model/Catalog Number: R-TRAK Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2845-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: Q6 Edge HD Model/Catalog Number: Q6 Edge HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2846-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum rehab Product Name: Q6 Edge 2.0 Model/Catalog Number: Q6 Edge 2.0 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2847-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: Quantum 1450 Model/Catalog Number: Quantum 1450 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2848-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: Q6 Edge 2.0 (ilevel) Model/Catalog Number: Q6 Edge 2.0 (ilevel) Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2849-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Model/Catalog Number: Q6 Edge 3 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2850-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: Q6 Edge 4 Model/Catalog Number: Q6 Edge 4 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2851-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: Quantum 4Front 2HD Model/Catalog Number: Quantum 4Front 2HD Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2852-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2853-2026

    Pride Mobility Products Corp.

    Brand Name: Quantum Rehab Product Name: Quantum Rival Model/Catalog Number: Quantum Rival Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

    Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this…

    Device · initiated 2026-06-25 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2890-2026

    Siemens Healthcare Diagnostics, Inc.

    Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic

    Reason: Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; Increased rate of reaction cuvette QC failures and Delayed…

    Device · initiated 2026-07-06 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2862-2026

    Stryker Corporation

    STRYKER AHTO TUBE SET PACKAGING, Catalog Number 0250070600

    Reason: There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use.

    Device · initiated 2026-06-08 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2863-2026

    Stryker Corporation

    STRYKER AHTO TUBE SET WITH TIP PACKAGING, Catalog Number 0250070620

    Reason: There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use.

    Device · initiated 2026-06-08 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2864-2026

    Stryker Corporation

    STRYKER AHTO DUAL SPIKE TUBE SET PACKAGING, Catalog Number 0250070640

    Reason: There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use.

    Device · initiated 2026-06-08 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2865-2026

    Stryker Corporation

    STRYKER PKG., SINGLE SPIKE TUBE SET, Catalog Number 0250070530

    Reason: There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use.

    Device · initiated 2026-06-08 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2866-2026

    Stryker Corporation

    STRYKER PKG., STRYKEFLOW SUCTION/IRRIGATOR PS, Catalog Number 0250070540

    Reason: There has been an increased amount in reported complaints consistently describing leaking from the handpiece during use.

    Device · initiated 2026-06-08 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2835-2026

    Therabody, Inc

    Charger for the Therabody Theragun, G4 PRO massage device, and products which include the 4th Generation Theragun PRO charger. G4 PRO Black US - Refurb., SKU TG02428-01 G4 PRO NA, SKU G4-PRO-PKG-NA G4-PRO-PKG-US, SKU G4 PRO US G4 PRO Black US - Refurb V, SKU TG0002428-1V10 Pro C…

    Reason: Charging cradle for the device may overheat while charging the device's removable battery pack, presenting risk of thermal event (fire). Firm has received reports alleging the product caught fire and other reports regarding visible smoke, melting, charger burn, burning odor, and…

    Device · initiated 2026-06-23 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2855-2026

    THERANICA BIOELECTRONICS LTD

    Nerivio Infinity. Model/Catalog Number: FGD000100. Software Version: FW Version 4.1.2.0, App version 7.28.0.11

    Reason: Potential for the batter to have a thermal-runway event during charging, resulting in smoke and odor.

    Device · initiated 2026-03-23 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2901-2026

    TISSIUM SA

    Brand Name: COAPTIUM Connect Product Name: COAPTIUM Connect Model/Catalog Number: NCOCO1510 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: N/A

    Reason: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device…

    Device · initiated 2026-06-15 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2902-2026

    TISSIUM SA

    Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO1575 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

    Reason: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device…

    Device · initiated 2026-06-15 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2903-2026

    TISSIUM SA

    Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

    Reason: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device…

    Device · initiated 2026-06-15 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2904-2026

    TISSIUM SA

    Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

    Reason: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device…

    Device · initiated 2026-06-15 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2905-2026

    TISSIUM SA

    Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO3010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

    Reason: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device…

    Device · initiated 2026-06-15 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2906-2026

    TISSIUM SA

    Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

    Reason: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device…

    Device · initiated 2026-06-15 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2907-2026

    TISSIUM SA

    Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO5010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

    Reason: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device…

    Device · initiated 2026-06-15 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2908-2026

    TISSIUM SA

    Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO6010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

    Reason: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device…

    Device · initiated 2026-06-15 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2883-2026

    United Orthopedic Corp.

    Brand Name: U2 Total Knee System Product Name: PSA Box Trial, L, #1-6 and PSA Box Trial, R, #1-6 Model/Catalog Number: 9323-5511 thru 9323-5516 and 9323-5521 thru 9323-5526 Software Version: N/A Product Description: Knee femur prosthesis trial, reusable Component: No, N/A

    Reason: Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.

    Device · initiated 2026-05-20 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2884-2026

    United Orthopedic Corp.

    Brand Name: USTAR II Knee System Product Name: Hinge Box Trial, L, #1-6 and Hinge Box Trial, R, #1-6 Model/Catalog Number: 9325-5711 thru 9325-5716 and 9325-5721 thru 9325-5726 Software Version: N/A Product Description: Knee femur prosthesis trial, reusable Component: No, N/A

    Reason: Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.

    Device · initiated 2026-05-20 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2885-2026

    United Orthopedic Corp.

    Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Model/Catalog Number: 9323-5333 and 9323-5334 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No

    Reason: Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.

    Device · initiated 2026-05-20 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2886-2026

    United Orthopedic Corp.

    Brand Name: U2 Total Knee System Product Name: Femoral valgus offset adaptor, 2mm, 4mm, 6mm, left and Femoral valgus offset adaptor, 2mm, 4mm, 6mm, right Model/Catalog Number: 9323-5335, 9323-5337, 9323-5339 and 9323-5336, 9323-5338, 9323-5340 Software Version: N/A Product Descr…

    Reason: Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.

    Device · initiated 2026-05-20 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2897-2026

    United Orthopedic Corp.

    Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 0mm and 4mm Model/Catalog Number: 9323-5310 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No, N/A

    Reason: Due to a design issue causing the incorrect engraved numbering on implant that could result in the incorrect orientation of the implant.

    Device · initiated 2026-03-19 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2898-2026

    United Orthopedic Corp.

    Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 2mm and 6mm Model/Catalog Number: 9323-5312 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No, N/A

    Reason: Due to a design issue causing the incorrect engraved numbering on implant that could result in the incorrect orientation of the implant.

    Device · initiated 2026-03-19 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • D-0745-2026

    Zydus Pharmaceuticals (USA) Inc

    Mirabegron Extended-Release Tablets, 25 mg, 30-count bottle, Rx only, Manufactured by: Zydus Lifesciences Ltd., Matoda, Ahmedabad, India, Distributed by: Zydus Pharmaceuticals (USA) Inc., Pennington, NJ 08534, NDC 70710-1159-3.

    Reason: CGMP: Due to Out of Specification (OOS) result for N-Nitroso Mirabegron impurity.

    Drug · initiated 2026-08-05 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
Class III3 recallsShow all
  • D-0740-2026

    Central Admixture Pharmacy Services Inc

    Total Parental Nutrition - Pediatric PN Patient-Specific TPN Bag, (patient specific), Compound Volume 416.8 mL per bag, Rx only, Single Dose Injection, Refrigerated Injection, Central Admixture Pharmacy Services, Inc., Atlanta, 1750 Corp Dr. Ste 725, Norcross, GA 30093 (844) 903…

    Reason: Incorrect Product Formulation

    Drug · initiated 2026-07-16 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2880-2026

    Siemens Healthcare Diagnostics, Inc.

    Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL

    Reason: Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post…

    Device · initiated 2026-06-26 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record
  • Z-2881-2026

    Siemens Healthcare Diagnostics, Inc.

    Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL

    Reason: out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and…

    Device · initiated 2026-06-26 · Enforcement Report 2026-08-12 · Ongoing

    openFDA record

New OAI classifications

1 inspection classified Official Action Indicated that FDAgent first recorded from FDA's inspection classification data this week, after its reads of that data in the 8 days before had not found them. Older inspections FDAgent adds to its archive are left out. Dates are inspection end dates. Inspection classifications last synced Nov 23, 2026, 04:25 UTC.

  • Inspection ended 2026-06-03

    Eurofins SF Analytical Laboratories, Inc.OAI

    Horsham, Pennsylvania, United States · Drugs · Drug Quality Assurance · recorded 2026-08-11

    FEI 3014704316

Get this page by email every Monday

A free weekly email with each week’s warning letters, new Form 483s, OAI classifications and Class I recalls, counted the same way as this page. Just your email address.

We’ll email once to confirm. Unsubscribe anytime. See last week’s issue →

About this data

  • Weeks are ISO weeks, Monday to Sunday, in UTC. Each section is dated by its own FDA date: posting date for warning letters, reading-room publish date for Form 483s (each inspection's 483 counts once, when FDA first listed it; FDA's re-posts and amended 483s with the FEI and record date of one it listed before are shown apart), Enforcement Report date for recalls, and the date published FDA prints beside each firm on an import alert. FDA publishes no date for an inspection classification, so an OAI classification counts as new on the day FDAgent first recorded the inspection, and only when FDAgent's complete reads of FDA's inspection data in the 8 days before had covered the inspection's end date without finding it. Older inspections FDAgent adds to its archive, by a backfill or a wider read, never count as new, however late FDA posted them. Inspections FDAgent recorded before it began keeping a record of its reads, in November 2026, count only when they ended in the 13 weeks before, and the Veterinary inspections of FDAgent's first read of them, that same month, never count.
  • Import alert entries come from FDA's import alert pages, where FDA lists firms by name and address, not by FEI. FDAgent links an entry to a facility only when FDA printed the firm's FEI, never by name. Earlier weeks list only entries still on FDA's pages, because FDA deletes an entry when it removes a firm, and FDAgent keeps removed entries only from the day it began recording them. FDAgent lists an entry here only when its name shows it is a firm, through a legal form such as Ltd, Inc or GmbH, a business word such as Laboratories or Trading, or a web address, or when FDA's inspection or drug registration records hold a firm under exactly that name; entries that may name a person are shown only to signed-in users.
  • Import refusals are not listed by week: FDA reports a refusal weeks after it happens, so a week's count would keep changing after the week closed. Each facility page lists its own refusals.
  • Warning letters are those issued by FDA's drug, biologics and device centers and their field and quality offices. Letters from FDA's food, tobacco, veterinary, cosmetics and dietary supplement offices are left out.
  • FDA posts only some Form 483s to its Electronic Reading Room, so a missing 483 does not mean none was issued.
  • FDA publishes an inspection's classification once it is final and updates its inspection data weekly. Source: FDA inspections FAQ.
  • Recalls come from openFDA, which republishes FDA's weekly Enforcement Report; a report's recalls appear here once openFDA publishes it.
  • FDAgent is independent of FDA. Every row links to its FDA source; check it before relying on a record.