Snap Diagnostics, LLC. Form 483, February 2026

FDA issued this Form 483 to Snap Diagnostics, LLC. in Vernon Hills, Illinois (FEI 3000215335) at the end of a device inspection that ended February 13, 2026. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 20, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Feb 13, 2026
Devices · Compliance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Oct 20, 2026
249 days after the inspection
Facility
FEI 3000215335
Vernon Hills, Illinois

FDA's 3 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. ISO 13485:2016 Clause 7.1

    Product realization risk management processes not documented

    One or more processes for risk management in product realization have not been adequately documented.

  2. ISO 13485:2016 Clause 8.3.1

    Procedure for nonconforming product not documented

    Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not adequately documented.

  3. ISO 13485:2016 Clause 8.5.2

    Corrective action not taken

    Corrective action was not adequately taken to eliminate the cause of nonconformities in order to prevent recurrence.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.