QPS, LLC Form 483, January 2026
FDA issued this Form 483 to QPS, LLC in Newark, Delaware (FEI 3003581184) at the end of a drug inspection that ended January 30, 2026. It lists 2 observations. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Jul 21, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Jan 30, 2026
- Drugs · Bioresearch Monitoring
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Jul 21, 2026
- 172 days after the inspection
The 2 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Failure to follow SOP TLM-005
Observation 1: The firm did not follow SOP TLM-005. Specially, during study ®X@>5 sample analysis, sample OV was initially assessed in run 206 and the concentration was found to be 438 ng/mL. Sample was re-assayed in error and then subsequently repeated three time. The repeating of the sample analysis three time is not consistent with SOP TLM-005 section 5.6.
- 2
Incomplete source records for freeze-thaw
Observation 2: The firm had source records with discrepant documentations that did not allow complete reconstruction of freeze-thaw cycle in method validation QPS-997-1704 data. Specifically, the aliquots used to evaluate 3 and 5 freeze-thaw cycle were not recorded in the source documentation.
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.