Nanovis LLC Form 483, March 2026
FDA issued this Form 483 to Nanovis LLC in Columbia City, Indiana (FEI 3009446038) at the end of a device inspection that ended March 19, 2026. FDA classified the inspection Official Action Indicated (OAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Nov 12, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Mar 19, 2026
- Devices · Project Evaluation: Devices
- FDA's outcome
- OAI
- Official Action Indicated
- Posted by FDA
- Nov 12, 2026
- 238 days after the inspection
FDA's 6 citations from this inspection
FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.
- ISO 13485:2016 Clause 4.1.4
Changes to the QMS not adequately controlled or evaluated
Changes to quality management system processes were not adequately controlled in accordance with requirements and evaluated for their impact on the quality management system and the medical devices produced.
- ISO 13485:2016 Clause 8.5.1
Changes to maintain QMS and device not identified and implemented
Changes necessary to ensure and maintain the continued suitability, adequacy, and effectiveness of the quality management system and medical device safety and performance were not adequately identified and implemented.
- ISO 13485:2016 Clause 7.5.6
Procedures for validation of processes lack required elements
Procedures for validation of processes did not adequately include statistical techniques with rationale for sample sizes and use of specific methods, procedures, and acceptance criteria.
- ISO 13485:2016 Clause 7.5.1
Production controls not adequate
Production controls do not include implementation of monitoring and measurement of process parameters and product characteristics.
- ISO 13485:2016 Clause 6.4.1
Work environment requirements and procedures not documented
Requirements for the work environment, and the procedures to monitor and control the work environment, were not adequately documented where conditions for the work environment can have an adverse effect on product quality.
- ISO 13485:2016 Clause 7.4.2
Purchasing information does not include all required elements
Purchasing information does not adequately describe or reference the product to be purchased, including requirements for product acceptance, procedures, processes, and equipment.
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.