Jardon Eye Prosthetics, Inc. Form 483, March 2026
FDA issued this Form 483 to Jardon Eye Prosthetics, Inc. in Southfield, Michigan (FEI 1824736) at the end of a device inspection that ended March 2, 2026. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Nov 17, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Mar 2, 2026
- Devices · Compliance: Devices
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Nov 17, 2026
- 260 days after the inspection
FDA's 8 citations from this inspection
FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.
- ISO 13485:2016 Clause 5.5.1
Management responsibilities and authorities not documented
Top management did not ensure that responsibilities and authorities are defined, documented, and communicated within the organization.
- ISO 13485:2016 Clause 8.3.1
Procedure for nonconforming product not documented
Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not documented.
- ISO 13485:2016 Clause 5.6.1
Management review procedures not documented
Procedures for management review have not been documented.
- ISO 13485:2016 Clause 4.2.4
Lack of document control
Documents required by the quality management system have not been adequately controlled.
- ISO 13485:2016 Clause 7.1
Product realization risk management processes not documented
One or more processes for risk management in product realization have not been documented.
- ISO 13485:2016 Clause 8.2.4
Internal audit procedures not documented
Procedures to describe the responsibilities and requirements for planning and conducting internal audits and recording and reporting audit results were not adequately documented.
- ISO 13485:2016 Clause 7.4.1
Criteria for the evaluation and selection of suppliers not established
Criteria for the evaluation and selection of suppliers were not established.
- ISO 13485:2016 Clause 6.3
Requirements for infrastructure maintenance activities not documented
Requirements for infrastructure maintenance activities that can affect product quality, including the interval of performing the maintenance activities, have not been adequately documented.
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.