Guardian Medical USA Inc. Form 483, June 2026

FDA issued this Form 483 to Guardian Medical USA Inc. in Swanton, Ohio (FEI 3016261131) at the end of a device inspection that ended June 2, 2026. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 20, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jun 2, 2026
Devices · Compliance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Nov 20, 2026
171 days after the inspection
Facility
FEI 3016261131
Swanton, Ohio

FDA's 6 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. ISO 13485:2016 Clause 7.3.8

    Design and development transfer procedures not documented

    Procedures for transfer of design and development outputs to manufacturing were not documented.

  2. ISO 13485:2016 Clause 7.4.1

    Criteria for the evaluation and selection of suppliers not established

    Criteria for the evaluation of suppliers were not adequately established.

  3. ISO 13485:2016 Clause 7.5.6

    Planned results not consistently met during validation of processes

    Validation of processes for production and service provision did not adequately demonstrate the ability of the processes to achieve planned results consistently.

  4. ISO 13485:2016 Clause 8.5.2

    Corrective action procedures not documented

    Corrective action procedures defining the requirements for reviewing nonconformities and evaluating the need for action to ensure that nonconformities do not recur have not been adequately documented.

  5. ISO 13485:2016 Clause 4.1.2

    Risk-based approach not applied to QMS processes

    A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.

  6. ISO 13485:2016 Clause 8.3.1

    Procedure for nonconforming product not documented

    Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not adequately documented.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.