Comfortland Medical, Inc. Form 483, June 2026

FDA issued this Form 483 to Comfortland Medical, Inc. in Mebane, North Carolina (FEI 3008724689) at the end of a device inspection that ended June 12, 2026. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 20, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jun 12, 2026
Devices · Compliance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Nov 20, 2026
161 days after the inspection
Facility
FEI 3008724689
Mebane, North Carolina

FDA's 8 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. ISO 13485:2016 Clause 8.2.2

    Complaint handling records not maintained

    Complaint handling records were not adequately maintained.

  2. ISO 13485:2016 Clause 7.5.1

    Production and service provision not adequate

    Production and service provision was not adequately planned, carried out, monitored, and controlled to ensure that product conforms to specification.

  3. Label to bear a unique device identifier

    The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

  4. ISO 13485:2016 Clause 4.1.5

    Control over outsourced processes not proportionate to risk

    Controls over outsourced processes that affect product conformity to requirements are not proportionate to the risk involved and the ability of the external party to meet purchasing requirements.

  5. Lack of Written MDR Procedures

    Written MDR procedures have not been developed, maintained and implemented.

  6. ISO 13485:2016 Clause 7.1

    Processes needed for product realization not planned and developed

    Processes needed for product realization were not planned and developed.

  7. ISO 13485:2016 Clause 4.1.1

    QMS not documented and maintained adequately

    The quality management system was not documented and effectiveness was not maintained in accordance with applicable regulatory requirements.

  8. ISO 13485:2016 Clause 7.3.10

    Design and development files not maintained

    A design and development file was not maintained for each medical device type or medical device family.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.