Skip to content

Clearwater Colon Hydrotherapy, Inc. Form 483, March 2026

FDA issued this Form 483 to Clearwater Colon Hydrotherapy, Inc. in Ocala, Florida (FEI 3000718678) at the end of a device inspection that ended March 26, 2026. FDA classified the inspection No Action Indicated (NAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 17, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Mar 26, 2026
Devices · Postmarket Assurance: Devices
FDA's outcome
NAI
No Action Indicated
Posted by FDA
Nov 17, 2026
236 days after the inspection
Facility
FEI 3000718678
Ocala, Florida

FDA's 6 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. ISO 13485:2016 Clause 7.2.2

    Review of requirements related to product not conducted

    Review of the requirements related to product was not conducted prior to committing to supply product to the customer.

  2. ISO 13485:2016 Clause 4.2.4

    Lack of document control

    Documents required by the quality management system have not been adequately controlled.

  3. ISO 13485:2016 Clause 7.4.1

    Supplier monitoring and re-evaluation not planned

    The re-evaluation of suppliers was not adequately planned.

  4. ISO 13485:2016 Clause 4.1.1

    QMS not documented and maintained adequately

    The quality management system was not documented in accordance with applicable regulatory requirements.

  5. ISO 13485:2016 Clause 7.5.6

    Records of validation of processes and necessary actions not maintained

    Records of the results and conclusion of validation of processes and necessary actions from the validation were not maintained.

  6. ISO 13485:2016 Clause 7.3.10

    Design and development file does not include or reference required records

    A design and development file does not adequately include or reference records generated to demonstrate conformity to the requirements for design and development.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.