Skip to content

Cedars-Sinai Medical Center Form 483, December 2015

FDA issued this Form 483 to Cedars-Sinai Medical Center in Los Angeles, California (FEI 3000717534) at the end of a device inspection that ended December 7, 2015. It lists 2 observations. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Dec 7, 2015
Devices · Postmarket Assurance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Jan 17, 2024
2,963 days after the inspection
Facility
FEI 3000717534
Los Angeles, California

The 2 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Missing written MDR procedure elements

    1. Written MDR procedures have not been developed. Specifically, your two paged written MDR procedure, entitled "Biomedical Equipment Safe Medical Device Act", dated 07/22/2010, does not address or refer to the following elements: A) The use of FDA Form 3500A or electronic equivalent for mandatory device adverse event reporting. B) The definition of a serious injury. C) Your user facility reporting number.

  2. 2

    Failure to report device adverse event

    2. The user facility did not submit FDA Form 3500A or electronic equivalent to the known device manufacturer within 10 working days of becoming aware of information that reasonably suggests that a device has [illegible] or contributed to a serious injury to a patient. Specifically, you submitted a voluntary FDA form 3500 for an event, dated [illegible], where a device malfunctioned and caused an intervention to prevent permanent impairment or damage to the patient. You did not identify this event with your user facility reporting number.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.