Bioplate Inc Form 483, June 2026
FDA issued this Form 483 to Bioplate Inc in Placentia, California (FEI 1000518793) at the end of a device inspection that ended June 1, 2026. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Nov 17, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Jun 1, 2026
- Devices · Compliance: Devices
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Nov 17, 2026
- 169 days after the inspection
FDA's 5 citations from this inspection
FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.
- ISO 13485:2016 Clause 7.1
Records of risk management activities in product realization not maintained
Records of risk management activities in product realization were not adequately maintained.
- ISO 13485:2016 Clause 7.1
Planning of product realization not consistent with other QMS processes
Planning of product realization was not consistent with the requirements of other quality management system processes.
- ISO 13485:2016 Clause 4.1.5
Inadequate monitoring and control over outsourced processes
Outsourced processes that affect product conformity to requirements were not adequately monitored and controlled.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for reviewing nonconformities, determining the causes of nonconformities and evaluating the need for action to ensure that nonconformities do not recur have not been adequately documented.
- ISO 13485:2016 Clause 8.2.1
Feedback process procedures not documented
Procedures for the feedback process have not been adequately documented to include provisions for gathering data from production and post-production activities.
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.