Skip to content

CARE LIVING DIAGNOSTICS INC

415 Oser Ave Ste S, Hauppauge, New York, United States

FEI 3031259044

FDA has one inspection on file for CARE LIVING DIAGNOSTICS INC in Hauppauge, New York (FEI 3031259044). It ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 5 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
1, in 2026
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Sep 16, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 8.3.1

      Procedure for nonconforming product not documented

      Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not documented.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    3. ISO 13485:2016 Clause 8.5.3

      Preventive action procedures not documented

      Preventive action procedures describing the requirements for determining potential nonconformities and their causes, evaluating the need for action to prevent occurrence of potential nonconformities, planning and documenting action needed and implementing such action, verifying that the action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of the preventive action taken have not been documented.

    4. ISO 13485:2016 Clause 8.2.2

      Complaint handling procedures not documented

      Procedures for timely complaint handling have not been documented.

    5. ISO 13485:2016 Clause 8.5.2

      Corrective action procedures not documented

      Corrective action procedures defining the requirements for reviewing nonconformities, determining the causes of nonconformities, evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been documented.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Sep 16, 2026. It asked for a Form 483.

  • Inspection ending Jul 10, 2026VAI

    1 request. In the inspection history above.

    1. Sep 16, 2026FDA-FOIA-2026-8952Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 1 section of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3031259044 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3031259044. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.