The University of Texas MD Anderson Cancer Center, Department of Radiation Physics, Division of Radiation Oncology
1515 Holcombe Blvd Unit 1636, Houston, Texas, United States
FEI 3030564015
FDA has one inspection on file for The University of Texas MD Anderson Cancer Center, Department of Radiation Physics, Division of Radiation Oncology in Houston, Texas (FEI 3030564015). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 9 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2026
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Sep 24, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.3.9
Requirements for design and development changes not met before implementation
Design and development changes were not adequately reviewed, verified, validated, and approved before implementation.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for reviewing the effectiveness of corrective action taken have not been documented.
- ISO 13485:2016 Clause 8.2.4
Internal audits not conducted to ensure QMS meets requirements and is effective
Internal audits have not been conducted at planned intervals to determine whether the QMS is effectively implemented and maintained.
- ISO 13485:2016 Clause 7.3.7
Design and development validation not performed as planned
Design and development validation was not adequately performed in accordance with planned and documented arrangements to ensure that the resulting product is capable of meeting the requirements for the specified application or intended use.
- ISO 13485:2016 Clause 7.4.3
Purchased product verification activities not established or implemented
The inspection or other activities necessary for ensuring that purchased product meets specified purchasing requirements were not adequately established and implemented.
Show 4 more citations
- ISO 13485:2016 Clause 7.5.6
Procedures for production/service provision software validation not documented
Procedures were not documented for the validation of the application of computer software used in production and service provision.
- ISO 13485:2016 Clause 7.3.6
Inadequate design and development verification for interfaced device
Design and development verification of medical devices intended to be connected to, or interfaced with, other medical devices did not include confirmation that the design outputs meet design inputs when so connected or interfaced.
- ISO 13485:2016 Clause 5.6.1
QMS not reviewed by top management at documented planned intervals
Top management did not adequately review the quality management system at documented planned intervals to ensure its continued suitability, adequacy, and effectiveness.
- ISO 13485:2016 Clause 8.2.2
Correction or corrective action resulting from complaint handling not documented
Correction or corrective action resulting from the complaint handling process was not adequately documented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Sep 24, 2026. Each one asked for a Form 483.
Inspection ending Jul 27, 2026VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 30, 2026. FDAgent holds no copy that FDA posted.
- Sep 24, 2026FDA-FOIA-2026-9251Form 483· 1 of 5 sitesOpen
- Sep 10, 2026FDA-FOIA-2026-8805Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3030564015 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3030564015. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.