FEI 3030492371
FDA has one inspection on file for The Side Lyer Company LLC in Orlando, Florida (FEI 3030492371). It ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 6 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 1, in 2026
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Jun 16, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 4.2.5
Records of conformity and QMS operations not maintained
Records to provide evidence of conformity to requirements and the effective operation of the quality management system have not been maintained.
- ISO 13485:2016 Clause 6.2(a)
Necessary personnel competence not determined
The necessary competence for personnel performing work that affects product quality has not been determined.
- ISO 13485:2016 Clause 7.4.1
Procedures for purchasing process not documented
Procedures to ensure that purchased product conforms to specified purchasing information have not been documented.
- ISO 13485:2016 Clause 7.3.9
Requirements for design and development changes not met before implementation
Design and development changes were not reviewed, verified, validated, and approved before implementation.
- ISO 13485:2016 Clause 7.1
Records of risk management activities in product realization not maintained
Records of risk management activities in product realization were not adequately maintained.
Show 1 more citation
- ISO 13485:2016 Clause 4.1.3
QMS process records not established or maintained
Records needed to demonstrate compliance with applicable regulatory requirements have not been established and maintained.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Jun 16, 2026. Each one asked for a Form 483.
Inspection ending Mar 12, 2026NAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Apr 15, 2026. FDAgent holds no copy that FDA posted.
- Jun 16, 2026FDA-FOIA-2026-5846Form 483· 1 of 4 sitesRecords released
- Mar 31, 2026FDA-FOIA-2026-3154Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3030492371 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3030492371. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.