EVOLUTION DEVICES, INC
1231 Stevenson St Ste 201, San Francisco, California, United States
FEI 3030418885
FDA has one inspection on file for EVOLUTION DEVICES, INC in San Francisco, California (FEI 3030418885). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 7 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2026
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Jul 7, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 8.5.2
Corrective action not taken
Corrective action was not adequately taken to eliminate the cause of nonconformities in order to prevent recurrence.
- ISO 13485:2016 Clause 8.2.2
Complaint handling procedures lack required elements
Procedures for timely complaint handling do not adequately include requirements and responsibilities for receiving and recording information, investigating complaints and determining the need to initiate corrections or corrective actions.
- ISO 13485:2016 Clause 7.6
Procedures for monitoring and measuring equipment requirements not documented
Procedures to ensure that monitoring and measurement can be carried out and are carried out in a manner that is consistent with the monitoring and measurement requirements were not documented.
- ISO 13485:2016 Clause 8.3.1
Procedure for nonconforming product not documented
Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not adequately documented.
- ISO 13485:2016 Clause 7.4.1
Procedures for purchasing process not documented
Procedures to ensure that purchased product conforms to specified purchasing information have not been adequately documented.
Show 2 more citations
- ISO 13485:2016 Clause 8.2.4
Internal audits not conducted to ensure QMS meets requirements and is effective
Internal audits have not been conducted at planned intervals to determine whether the QMS.
- ISO 13485:2016 Clause 7.1
Product realization risk management processes not documented
One or more processes for risk management in product realization have not been documented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Jul 7, 2026. Each one asked for a Form 483.
Inspection ending Apr 15, 2026VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 24, 2026. FDAgent holds no copy that FDA posted.
- Jul 7, 2026FDA-FOIA-2026-6454Form 483· 1 of 4 sitesRecords released
- Jun 18, 2026FDA-FOIA-2026-5922Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3030418885 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3030418885. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.