FDA has one inspection on file for Reprolife America, LLC in Kingwood, Texas (FEI 3018010054). It ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 10 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
1, in 2026
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Jun 1, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.3.10

      Design and development file does not include or reference required records

      A design and development file does not include or reference records generated to demonstrate conformity to the requirements for design and development.

    2. ISO 13485:2016 Clause 7.5.9.1

      Procedures for traceability were not documented

      Procedures for traceability were not documented.

    3. ISO 13485:2016 Clause 7.5.6

      Process for production and service provision not validated

      Processes for production and service provision where the resulting output cannot be, or is not, verified by subsequent monitoring or measurement were not validated.

    4. ISO 13485:2016 Clause 6.4.1

      Work environment requirements and procedures not documented

      Requirements for the work environment, and the procedures to monitor and control the work environment, were not documented where conditions for the work environment can have an adverse effect on product quality.

    5. ISO 13485:2016 Clause 8.2.2

      Complaint handling procedures not documented

      Procedures for timely complaint handling have not been documented.

    Show 5 more citations
    1. ISO 13485:2016 Clause 8.5.2

      Corrective action procedures not documented

      Corrective action procedures defining the requirements for reviewing nonconformities, determining the causes of nonconformities, evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been documented.

    2. ISO 13485:2016 Clause 4.1.1

      QMS not documented and maintained adequately

      The quality management system was not documented in accordance with applicable regulatory requirements.

    3. ISO 13485:2016 Clause 7.1

      Product realization risk management processes not documented

      One or more processes for risk management in product realization have not been documented.

    4. ISO 13485:2016 Clause 7.3.9

      Significance of design and development changes not determined

      The significance of the design and development change to function, performance, safety and regulatory requirements for the medical device and its intended use was not determined.

    5. ISO 13485:2016 Clause 4.2.3

      Medical device file not established and maintained

      Medical device files have not been established and maintained.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Jun 1, 2026. It asked for a Form 483.

  • Inspection ending Mar 20, 2026VAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 14, 2026. FDAgent holds no copy that FDA posted.

    1. Jun 1, 2026FDA-FOIA-2026-5409Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3018010054 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3018010054. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.