Skip to content

Emerge Technologies Inc

955 Freeport Pkwy Ste 100, Coppell, Texas, United States

FEI 3016750352

FDA has one inspection on file for Emerge Technologies Inc in Coppell, Texas (FEI 3016750352). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 12 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2026
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Jun 24, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 8.5.3

      Preventive action procedures not documented

      Preventive action procedures describing the requirements for determining potential nonconformities and their causes, evaluating the need for action to prevent occurrence of potential nonconformities, planning and documenting action needed and implementing such action, verifying that the action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of the preventive action taken have not been documented.

    2. ISO 13485:2016 Clause 7.4.1

      Procedures for purchasing process not documented

      Procedures to ensure that purchased product conforms to specified purchasing information have not been adequately documented.

    3. ISO 13485:2016 Clause 6.2

      Process(es) for personnel competence, training and awareness not documented

      Process(es) for establishing competence, providing needed training and ensuring personnel awareness have not been documented.

    4. Label to bear a unique device identifier

      The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

    5. ISO 13485:2016 Clause 5.6.1

      Management review procedures not documented

      Procedures for management review have not been documented.

    Show 7 more citations
    1. ISO 13485:2016 Clause 8.5.2

      Corrective action procedures not documented

      Corrective action procedures defining the requirements for reviewing nonconformities, determining the causes of nonconformities, evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been documented.

    2. ISO 13485:2016 Clause 8.2.4

      Internal audit procedures not documented

      Procedures to describe the responsibilities and requirements for planning and conducting internal audits and recording and reporting audit results were not documented.

    3. ISO 13485:2016 Clause 4.2.1

      Quality management system documentation not adequate

      The quality management system documentation does not include a quality manual.

    4. ISO 13485:2016 Clause 4.1.2

      Risk-based approach not applied to QMS processes

      A risk-based approach to the control of the appropriate processes needed for the quality management system has not been applied.

    5. ISO 13485:2016 Clause 8.2.2

      Complaint handling procedures not documented

      Procedures for timely complaint handling have not been documented.

    6. ISO 13485:2016 Clause 4.1.6

      QMS software not validated prior to use or changes

      Quality management system software was not validated prior to its initial use.

    7. Lack of Written MDR Procedures

      Written MDR procedures have not been developed, maintained and implemented.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Jun 24, 2026. It asked for a Form 483.

  • Inspection ending Apr 24, 2026OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 10, 2026. FDAgent holds no copy that FDA posted.

    1. Jun 24, 2026FDA-FOIA-2026-6048Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 2 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

  • 21 CFR 801.20Label to bear a unique device identifier1 of 1
  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3016750352 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3016750352. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.