Biomat USA Inc
2950 S John Redditt Dr Ste 105, Lufkin, Texas, United States
FEI 3014028128
FDA has one inspection on file for Biomat USA Inc in Lufkin, Texas (FEI 3014028128). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA posted the Form 483 from that inspection. FDAgent read its 4 observations from the PDF; they are listed below. Three FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2026
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 3 requests, latest Sep 16, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsPosted 483 · 8 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- Observation 1
Educational material not presented understandably
Failure to present educational material to the donor in an appropriate form and in a manner designed to be understood by the donor. (b) (6), (b) (7)(C) donor [illegible]. The donor's profile in the blood establishment system notifies your (b) (6), (b) (7)(C) employee that the donor must be [illegible]. Your employee failed to (b) (6), (b) (7)(C) to the donor since education materials relating to risk factors associated with relevant transfusion-transmitted infections and risks and hazards associated with donating. Your employee asked the donor if they read the educational materials and medication deferral list and marked [illegible] on the questionnaire. The SOP "Accommodating Donors with Disabilities" states, for (b) (4) donors, "A staff member will ask all screening questions ... All Job Aids will be displayed in (b) (4). The staff member will (b) (4) (b) (4)" .. all documents normally presented to be read by a prospective or repeat donor and understood by them as a part of the initial and subsequent screening process must be read in their entirety to the donor by a center employee." (b) (6), (b) (7)(C) [illegible] This donor has been donating since [illegible] and has made [illegible] donations.
- Observation 2
Incomplete investigation of duplicate donor accounts
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy. Specifically, your firm opened a deviation 2196858 on 11/19/22 which identified a donor with multiple accounts under (b) (6), (b) (7)(C). This issue was discussed during the previous inspection. This incident was reported to the FDA (BPD 851423) and products were identified as effected. The donor's profiles were merged on 12/23/22 after the donor provided proof of identification. Your Regional Quality Manager and Quality Systems Manager were unable to provide documentation related to the root cause of the incident and additional corrective action related to prevention of reoccurrence. Additionally, in the report to FDA, it was stated the donor was permanently deferred. However, records indicate the donor was temporarily deferred until they could provide proof of identification. The donor (b) (6), (b) (7)(C) continued to donate until [illegible] and has a total of [illegible] donations.
Show 6 more observations
- Observation 3
Failure to maintain accurate donor records
Failure to maintain donor, processing and general records. Specifically, you failed to maintain accurate records during the eligibility process and when treating donor adverse events. 1. When documenting adverse events, your employees fail to document treatment given to donors. On 4/9/26, your donor center specialist performed a venipuncture where blood spilled and (b) (4), (b) (6), (b) (7)(C) (b) (6), (b) (7)(C). Your firm classified this as a minor donor adverse event. The documentation of the event did not include the treatment given to the donor. The SOP "Event Types and Information to Record" states, for a (b) (4) Uncomplicated "Document symptoms, actions taken to resolve the adverse event, and if symptoms resolved." The following are four examples of minor donor adverse event records where your employees (b) (6), (b) (7)(C) failed to include treatment given to the donor. Donor ID: [illegible], donation on [illegible]; donor ID: [illegible], donation on [illegible] and donor ID: [illegible] on [illegible] 2. On 4/9/26 your donor screening technician failed to document a consultation with your center (b) (6), (b) (7)(C) (b) (6), (b) (7)(C) After consultation with the medical specialist regarding a donor's pulse (ID [illegible] center medical specialist, the donor was cleared to donate. The SOP "Donor Requirement - Blood Pressure and Pulse" states, "If the medical staff determines that the donor does not have (b) (4) pulse, then documentation of this evaluation must be included in the donor's record." 3. When drawing blood during the screening process, (b) (4) capillary tubes were filled from the (b) (4) capillary finger, and it was not documented. Your donor screening technician filled (b) (4) (b) (6), (b) (7)(C) tubes from (b) (4) finger stick (donor ID: [illegible]. Furthermore, your SOP does not instruct employees to document when (b) (4) capillary tubes are filled from the (b) (4) finger stick, and it only instructs your employees to document when (b) (4) finger sticks are performed. (b) (4) (b) (4) You collected [illegible] units of plasma in 2026 and [illegible] units of plasma in 2025.
- Observation 4
SOPs for collection and storage not followed
Written standard operating procedures including all steps to be followed in the collection and storage of blood and blood components for [illegible] were not always followed. Specifically, the following are examples where your employees did not follow standard operating procedures (SOPs). 1. Your employees filled the capillary tubes, used to determine hematocrit and total protein, (b) (4) (b) (4). The SOP "Donor Requirement - Hematocrit" states, "(b) (4) (b) (4)" The operator's manual for the (b) (4) states, "Fill a capillary tube (b) (4)" For (b) (6), (b) (7)(C) out of [illegible] donor screenings observed, donor screening technicians filled the capillary tubes (b) (4) (b) (6), (b) (7)(C) (b) (6), (b) (7)(C) 2. Your donor screening technician failed to clean the (b) (4) during performance of (b) (4) quality control on 4/9/26. The SOP "Vital Sign Equipment Standardization" states, "Follow the below steps to complete the quality control:" (b) (4) "Clean the unit per instructions in the Operator's manual using (b) (4) (b) (4) 3. Your employees are not ensuring source plasma products in the freezers are set up according to your procedures. Your procedures state products should be (b) (4) I observed product was stored (b) (4) and In freezer (b) (4). In freezer I observed (b) (4) and (b) (4) Freezer product was stored (b) (4) (b) (4) [illegible] holds products and (b) (4), and freezer holds product that are (b) (4) (b) (4) products and (b) (4) (b) (4) (b) (4) units of plasma in 2025. You collected (b) (4) units of plasma in 2026 and [illegible]
Form and manner of educational material
Failure to present educational material to the donor in an appropriate form and in a manner designed to be understood by the donor.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection and storage of blood and blood components for were not always followed.
Required records
Failure to maintain donor, processing and general records.
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Sep 16, 2026. Each one asked for a Form 483.
Inspection ending Apr 14, 2026VAI
3 requests. In the inspection history above. FDA has posted its Form 483.
- Sep 16, 2026FDA-FOIA-2026-8969Form 483Records released
- Aug 4, 2026FDA-FOIA-2026-7477Form 483· 1 of 3 sitesRecords released
- Jun 10, 2026FDA-FOIA-2026-5685Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 3 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 606.100Standard operating procedures1 of 1
- 21 CFR 606.160Records1 of 1
- 21 CFR 630.10General donor eligibility requirements1 of 1
Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.
Other FEIs recorded as Biomat USA Inc
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Biomat USA IncFEI 3006118912Kingsport, Tennessee, United States
Last inspected 2025-03-05NAI
Biomat USA IncFEI 3012230191Greenville, North Carolina, United States
Last inspected 2025-02-07NAI
Biomat USA IncFEI 2970021Reno, Nevada, United States
Last inspected 2025-01-30NAI
Biomat USA IncFEI 3011871534Kissimmee, Florida, United States
Last inspected 2025-01-16NAI
Biomat USA IncFEI 3003824349Texarkana, Texas, United States
Last inspected 2024-12-06NAI
Biomat USA IncFEI 3003853409Portsmouth, Virginia, United States
Last inspected 2024-11-15NAI
Biomat USA IncFEI 2077841Glendale, Arizona, United States
Last inspected 2024-11-01NAI
Biomat USA IncFEI 3006215451Benson, North Carolina, United States
Last inspected 2024-10-31NAI
Biomat USA IncFEI 3007011613New Orleans, Louisiana, United States
Last inspected 2024-10-31NAI
Biomat USA IncFEI 3011626200Austin, Texas, United States
Last inspected 2024-09-20NAI
Not every site is listed: these are the most recently inspected sites found under this name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3014028128 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3014028128. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.