Skip to content

ALIEN MILLING TECHNOLOGIES

614 S Glendora Ave, Glendora, California, United States

FEI 3013376795

FDA has one inspection on file for ALIEN MILLING TECHNOLOGIES in Glendora, California (FEI 3013376795). It ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 7 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
1, in 2026
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Aug 5, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.4.1

      Procedures for purchasing process not documented

      Procedures to ensure that purchased product conforms to specified purchasing information have not been adequately documented.

    2. ISO 13485:2016 Clause 8.5.2

      Corrective action not taken

      Corrective action was not adequately taken to eliminate the cause of nonconformities in order to prevent recurrence.

    3. ISO 13485:2016 Clause 7.1

      Processes needed for product realization not planned and developed

      Processes needed for product realization were not planned and developed.

    4. ISO 13485:2016 Clause 4.2.4

      Lack of document control

      Documents required by the quality management system have not been controlled.

    5. ISO 13485:2016 Clause 8.3.1

      Nonconforming product not adequately evaluated

      Nonconforming product was not evaluated to determine the need for an investigation and notification of any external party responsible for the nonconformity.

    Show 2 more citations
    1. ISO 13485:2016 Clause 7.5.6

      Process for production and service provision not validated

      Processes for production and service provision where the resulting output cannot be, or is not, verified by subsequent monitoring or measurement were not adequately validated.

    2. ISO 13485:2016 Clause 5.1

      Lack of evidence for top management's commitment to QMS

      Top management did not adequately provide evidence of its commitment to the development and implementation of the quality management system and maintenance of its effectiveness by.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Aug 5, 2026. Each one asked for a Form 483.

  • Inspection ending Jun 25, 2026VAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 27, 2026. FDAgent holds no copy that FDA posted.

    1. Aug 5, 2026FDA-FOIA-2026-7542Form 483· 1 of 5 sitesRecords released
    2. Jul 31, 2026FDA-FOIA-2026-7330Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3013376795 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3013376795. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.