Advanced Magnetic Research Institute International LLC
6230 E Tropical Pkwy, Las Vegas, Nevada, United States
FEI 3010216897
FDA has one inspection on file for Advanced Magnetic Research Institute International LLC in Las Vegas, Nevada (FEI 3010216897). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 10 citations from this inspection. They are listed below.
No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 1, in 2013
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sponsor final report for non-significant risk study
For an investigation subject to the abbreviated requirements, a final report was not submitted to all reviewing IRBs within six months after termination and completion of an investigational device study.
Sponsor records re QSR/GMP & non-significant risk studies
For an investigation subject to the abbreviated requirements, records that include a statement of the extent to which the regulations covering current good manufacturing practice would be followed in manufacturing the device reports are not all complete.
Sponsor progress reports for non-significant risk study
For an investigation subject to the abbreviated requirements, progress reports were not submitted to all reviewing IRBs at least annually.
Sponsor evaluation rpt not timely, distributed
Reports of the results of evaluation of unanticipated adverse device effects were not all submitted to FDA.
Sponsor notif. of rqst. to return/repair/dispose not made
Notification was not sent to FDA of a request that an investigator return and dispose of any units of a device.
Show 5 more citations
Abbreviated requirements / non-significant risk study
A clinical investigation that was determined to be a non-significant risk device study does not meet the abbreviated requirements for investigational device exemptions (IDEs).
Sponsor records of non-significant risk study IRBs
For an investigation subject to the abbreviated requirements, records that include the name and address of each IRB that has reviewed the investigation are not all complete and current.
Promotion or test marketing
An investigational device was promoted prior to FDA approval/clearance.
Sponsor records re device use in non-significant risk study
For an investigation subject to the abbreviated requirements, records which include the name and intended use of the device are not all accurate, complete, and current.
Availability of records of non-significant risk studies
For an investigation subject to the abbreviated requirements, records are not available for FDA inspection and copying.
Regulations FDA cited here
The 4 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 812.2Applicability1 of 1
- 21 CFR 812.7Prohibition of promotion and other practices1 of 1
- 21 CFR 812.140Records1 of 1
- 21 CFR 812.150Reports1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3010216897 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3010216897. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.