Skip to content

Beijing Laserworld Int'l Import & Export, Co., Ltd

Beiwu Village, Shunyi District, Beijing, China

FEI 3006762146

FDA has 2 inspections on file for Beijing Laserworld Int'l Import & Export, Co., Ltd in Beijing, China (FEI 3006762146), from 2010 to 2011. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection. They are listed below.

FDA’s Import Alert 95-05 Red List names this firm at this address. FDA printed no FEI on the entry. No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
2, 2010–2011
0 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
1 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
95-05Red List
Matched on FDA’s firm name and street address

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Beam attenuator

      A laser system classified as a Class IV laser system was not provided with one or more permanently attached means, other than laser energy source switches, electrical supply main connectors, or the key-actuated master control, capable of preventing access by any part of the human body to all laser and collateral radiation in excess of the accessible emission limits of Class I and table VI.

    2. User info - list controls, etc.

      The manufacturer did not provide or cause to be provided a listing of the warning: "Caution --- use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.".

    3. Legible reproductions

      Legible reproductions of all labels and hazard warnings required by the regulations were not provided with the laser product.

    4. Protective housing

      Each laser product is not provided with a protective housing that prevents human access during operations to laser and collateral radiation that exceeds the limits of Class I and table VI, respectively, wherever and whenever such human access is not necessary for the product to perform its intended function(s).

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety ActCitations not published

Regulations FDA cited here

The 1 section of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.

Import alerts

FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.

  • 95-05Red ListDetention Without Physical Examination of Electronic Products that Fail to Comply with Performance Standards or to Have Required CertificationImport alert pageAlert on fda.gov

    Date published 2012-10-04

    FDA’s entry
    Beijing Laserworld Int'l Import & Export, Co., Ltd · Beiwu Village, Shunyi District, Beijing, Beijing CHINA
    FDA’s inspection record for FEI 3006762146
    Beijing Laserworld Int'l Import & Export, Co., Ltd · Beiwu Village, Shunyi District, Beijing, China

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    2 products

    • 95 R - - FELaser Cutter, Material Processing Laser Products
    • 95 R - - HKLaser marker or engraver

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3006762146 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3006762146. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.