C.A.T. GmbH and Co. KG
Heerweg 10, Tubingen, Germany
FEI 3004422881DUNS 324868512
FDA has 4 inspections on file for C.A.T. GmbH and Co. KG in Tubingen, Germany (FEI 3004422881), from 2012 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.
No warning letter, court action or import refusal is filed under this FEI.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for analysis, through Dec 31, 2027. Its registrant is C.A.T. GmbH & Co. Chromatographie und Analysentechnik KG (D-U-N-S 324868512).
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2012–2024
- 0 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 1 of 4
- Form 483s posted by FDA
- ↗
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 1 observation
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Insufficient chromatography system suitability
OBSERVATION 1 The establishment of test procedures and laboratory control mechanisms including any changes thereto, are not drafted by the appropriate organizational unit and reviewed and approved by the quality control unit. Specifically, USP <621> Chromatography states that system suitability tests are an integral part of gas and liquid chromatographic methods which verify that the system is adequate for the intended analysis. It states that replicate injections of standard solutions are compared to ascertain whether precision requirements are met. Additionally, it states that system suitability ascertains final operating system effectiveness, and that additional injections of standard solutions are required at appropriate times to demonstrate system suitability throughout the chromatography run. Using chromatographic methods, your firm analyzed residual solvents and/or the optical purity of [illegible] of multiple lots of the Active Pharmaceutical Ingredient (API) for the drug product [illegible] including lots [illegible]. However, your firm did not establish system suitability through multiple consecutive injections of the appropriate standard solutions. According to the CEO and Owner, your firm currently performs system suitability by analyzing injection of the appropriate standard solutions [illegible]. Additionally, your firm did not confirm the effectiveness of the final operating system through injecting standard solutions at appropriate intervals and after the final API sample injection.
FDA drug establishment registration
FDA registration search- Registered operations
- Analysis
- Registration runs through
- Dec 31, 2027
- Registrant
- C.A.T. GmbH & Co. Chromatographie und Analysentechnik KGD-U-N-S 324868512
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004422881 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004422881. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.