Alps Pharmaceutical Industrial Co., Ltd.
Japan
FEI 3002806352
FDA has one inspection on file for Alps Pharmaceutical Industrial Co., Ltd. in Japan (FEI 3002806352). It ended on ; FDA has not posted its classification yet.
FDA posted the Form 483 from that inspection. FDAgent read its 2 observations from the PDF; they are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- Inspections on file
- 1, in 2025
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 1 request, latest Nov 22, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- Not classified yetClassification not posted yetPosted 483 · 2 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
HPLC instrument time not synchronized
QC laboratory established written procedure for synchronization of laboratory instruments time with the actual time was not followed. For example, QC laboratory SOP: 013-41, Quality Control Procedure, Effective Date: 2025-05-23, Version 41, states that QC laboratory instruments time shall be synchronized within plus or minus (b)(4) of the actual time. The hardware time display of the HPLC was not synchronized with the network LabSolution software. On 10/16/2025 during the walk through inspection of the QC laboratory the hardware time display of #HPLC100-6-34 displayed the time difference of (b)(4) and #HPLC100-6-40 displayed the time difference of (b)(4) from the actual time which did not meet the requirement as per the established written procedure. HPLC #HPLC100-6-34 is used for release test of (b)(4) and HPLC #HPLC100-6-40 is used for release testing of (b)(4) API that are either shipped directly or indirect to the US customers.
- Observation 2
No incubator cleaning or media labeling
Microbiological laboratory lacked established written procedure for cleaning of incubators and labeling of prepared media, for example, A. There is no established written procedure for cleaning of incubators in which incubator #H100-25-7 inside cabinet seal door was observed containing mold and mildews on 10/15/2025 during the walk through inspection. Incubator #H100-25-7 is used to incubate media plates for microbiological testing of finished APIs that are shipped to the US. B. There is no established written procedure for proper labeling media (b)(4) plates. The current practice of labeling prepared media (b)(4) plates does not contain the initial of the analyst, the date of inoculation, and identification of the sample that can be tracked to the analytical worksheet.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Nov 22, 2025. It asked for a Form 483.
Inspection date not given
1 request. FDA released records in answer to a request for its Form 483 on Dec 11, 2025. FDAgent holds no copy that FDA posted.
- Nov 22, 2025FDA-FOIA-2025-13231Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Other FEIs recorded as Alps Pharmaceutical Industrial Co., Ltd.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Alps Pharmaceutical Industrial Co., Ltd.FEI 1073Hida, Japan
Last inspected 2025-10-17VAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002806352 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002806352. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.