Governor Juan F. Luis Hospital & Medical Center
4007 Estate Diamond Ruby, Christiansted, Virgin Islands, United States
FEI 2673291
FDA has 6 inspections on file for Governor Juan F. Luis Hospital & Medical Center in Christiansted, Virgin Islands (FEI 2673291), from 2010 to 2016. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 606, the manufacturing rules for blood and blood components. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2010–2016
- 0 OAI · 6 VAI · 0 NAI
- Inspections with 483 observations shown
- 6 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.
Required records
Failure to maintain storage and distribution and general records.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Weight
Donor did not weigh 110 pounds or more on the date of donation.
Controlling storage temperatures
The standard operating procedure fails to include a written description of the storage temperatures and methods of controlling storage temperatures for all blood products and reagents
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.
Required records
Failure to maintain storage and distribution and general records.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection and storage of blood and blood components for homologous transfusion are not always maintained and followed.
Required records
Failure to maintain donor, storage and distribution, and general records.
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.
Person performing, test results, interpretation
Records fail to identify the person performing the work, show test results, include interpretation of the results, and be as detailed as necessary so as to provide a complete history of the work performed.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Concurrent documentation
Records are not concurrently maintained with the performance of each significant step in the collection of each unit of blood and blood components so that all steps can be clearly traced.
Thorough investigations
Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Legibility and indelibility
Records are illegible.
Personnel capabilities
Failure to assure that personnel have capabilities commensurate with and a thorough understanding of the operations which they perform.
Establish, maintain and follow manufacturing SOPs
Written standard operating procedures including all steps to be followed in the collection and processing of blood and blood components for homologous transfusion and further manufacturing purposes are not always followed.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Person performing, test results, interpretation
Records fail to identify the person performing the work, include dates of the various entries, and be as detailed as necessary so as to provide a complete history of the work performed.
Qualifications of responsible personnel
The personnel responsible for the collection of blood or blood components are not adequate in educational background and training and experience, including professional training as necessary to assure competent performance of their assigned functions, and to ensure that the final product has the safety, purity, potency, identity and effectiveness it purports or is represented to possess.
Legibility and indelibility
Records are illegible.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 606.160Records6 of 6
- 21 CFR 606.100Standard operating procedures5 of 6
- 21 CFR 600.10Personnel1 of 6
- 21 CFR 606.20Personnel1 of 6
- 21 CFR 640.631 of 6
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2673291 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2673291. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.