Utah Medical Products, Inc
7043 S 300 W, Midvale, Utah, United States
FEI 1718873
FDA has 2 inspections on file for Utah Medical Products, Inc in Midvale, Utah (FEI 1718873), from 2010 to 2014. The latest ended on and was classified NAI, no action indicated. FDA has also posted a Form 483 dated Jul 23, 2015 that FDAgent has not matched to an inspection on the Data Dashboard.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA posted the Form 483 from that inspection. FDAgent read its 5 observations from the PDF; they are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 2, 2010–2014
- 0 OAI · 1 VAI · 1 NAI
- Inspections with 483 observations shown
- 1 of 2
- 4 posted 483s not matched to an inspection
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- NAINo objectionable conditions that need follow-upDevices · Compliance: Devices5 citations published
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
CAPA system deficiencies
1 Review of the firm's corrective and preventive action system revealed: Ia) Corrective and Preventive Action procedure, TD-QA-0 dataco identify existing and potential product and quality problems. b) Not all quality data are being analyzed to identify existing and potential product and quality problems. For example: in-process rejects and MDRs. c) There are no corrective and preventive actions taken for \./ "/ "-./ x x·· /\ _//~ /"" / " 'v.:. ""><.:.. )..,~ "<. >< '-.,.c '><::"
- Observation 2
DHRs and testing documentation issues
2 Review of the firm's Device History Records (DHRs) for the Intran Plus Catheters, ICP-400 revealed a) Manufacturing Procedure "Intran Plus and IUP-300 Final Tester" does not all approved design specifications prior w acceptance. Device Master Record specifications y..;: :x._ x releaseor'fmisheddevices in b) There are no documented evidence that the following required tests are being performed: >~ ':><-- .><.. >< .><::. c) Not all rejects are identified and documented. The devices failed during j Teste:- Summary Report.docs not indicate final disposition.
Show observations 4 to 6
- Observation 4
Nonconforming material procedure gaps
4. Review of the firm's Non-Conforming Material system and a) Non-Conforming Materials procedure X does b) The Non-Conforming Materials procedure, -..<: X - Rework and Use As Is dispositions do not always include an assessment the quality of the final product ----;-<:: •.-<:.. - Use As Is dispositions do not always include the justification for the use
- Observation 5
No FDA certification for e-signatures
5. The firm has yet to certify to FDA that the electronic signatures eouivaknt of traditional handwritten signatures. Electronic inspection systems.
- Observation 6
Electronic records/signatures procedures missing
6. For the electronic records and signatures, there are no procedures addressing the following: a) Validation of systems to ensure accuracy, reliability, consistent intended per: recorded records. b) The ability to generate accurate and complete copies of c) Protection of records throughout the record retention period. d) Limiting system access. e) And the systems can create an audit trail that is computer-generated, time stamped use of operator/one and actions.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1718873 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1718873. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.