2026-W35 · 2026-08-24 to 2026-08-30
FDA enforcement, week of 2026-08-24
In 2026-W35 (2026-08-24 to 2026-08-30), FDA posted 1 warning letter from its drug, biologics and device offices, and listed no new Form 483s in its Electronic Reading Room. FDA's Enforcement Report added 60 drug and device recalls, 6 of them Class I. FDAgent recorded 2 new Official Action Indicated (OAI) classifications in FDA's inspection data. FDA published 5 firm entries on its import alert lists, 4 of them on a Red List.
- Warning letters posted
- 1
- Posted 2026-08-25
- Form 483s listed
- 0
- In the reading room
- Recalls
- 60
- 6 Class I · 52 II · 2 III
- New OAI classifications
- 2
- Newly listed by FDA
Ask FDAgent about this week
FDAgent reads the warning letters FDA posted this week and answers in plain English, with links to the letters.
Warning letters
1 letter FDA posted on its warning letters page this week, from its drug, biologics and device offices. Each letter's issue date can be weeks before its posting date. Warning letters last synced Nov 23, 2026, 04:17 UTC.
Unapproved New Drugs/Unlicensed Biological Product Violations
CBER · issued 2026-08-14 · posted 2026-08-25
An FEI links to its facility page only when the letter names exactly one FEI and FDAgent holds FDA records for it.
Import alert entries
5 firm entries FDA published on its import alert pages this week, dated by each entry's date published: 4 on a Red List, 1 on a Green List. FDA lists firms by name and address; FDAgent links an entry to a facility only when FDA printed the firm's FEI. 1 entry whose name may be a person's rather than a firm's is not listed or counted here; signed-in users can see them in FDAgent. Import alerts last synced Nov 23, 2026, 04:17 UTC.
66-79Red List2 entriesAlert on fda.govShow allDetention Without Physical Examination of Drugs From Foreign Establishments Refusing FDA InspectionAlert on fda.gov
- Published 2026-08-26
Chongqing Qiancan Technology Co., Ltd.
CHINA
- Published 2026-08-25
Tianjin Healthcare Medical Technology Co., Ltd
CHINA
76-01Green List1 entryAlert on fda.govShow allDetention Without Physical Examination Of Medical Instruments from PakistanAlert on fda.gov
- Published 2026-08-25
Golden Surgical International
PAKISTAN
95-05Red List2 entriesAlert on fda.govShow allDetention Without Physical Examination of Electronic Products that Fail to Comply with Performance Standards or to Have Required CertificationAlert on fda.gov
- Published 2026-08-26
Shenzhen Wanray Electronic Technology Co., Ltd.
CHINA
- Published 2026-08-24
Hubei Fithope Technology
CHINA
Form 483s listed in FDA's Electronic Reading Room
0 Form 483s FDA listed for the first time in its Electronic Reading Room this week, dated by the publish date FDA gives each. FDA sometimes lists a 483 days after that date, so a week can still gain 483s after it ends. FDA posts only some 483s, often months after the inspection. Reading room last synced Nov 23, 2026, 04:33 UTC.
FDA listed no new Form 483s in the reading room this week.
Recalls by class
Drug and device recalls in FDA's Enforcement Report dated this week, from openFDA, which FDA last updated Nov 11, 2026.
Class I6 recallsShow all
- D-0800-2026openFDA record
American Regent, Inc.
Papaverine HCl Injection, USP, 60 mg/2 mL (30 mg/mL), packaged in a) 2mL Single-Dose Vial (NDC 0517-4002-01), and b) 25x2mL Single-Dose Vials (NDC 0517-4002-25), Rx Only, AMERICAN REGENT, INC., SHIRLEY, NY 11967.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- Z-2872-2026openFDA record
Bard Access Systems, Inc.
BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Drive…
Reason: After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-o…
Device · initiated 2026-07-30 · Enforcement Report 2026-08-26 · Ongoing
- Z-2979-2026openFDA record
BMC Medical Co., Ltd.
BMC/REACTHEALTH Luna G3 APAP REF LG 3600
Reason: Retroactive reporting for 2024 rework conducted at the 3B/React Health warehouse
Device · initiated 2026-07-15 · Enforcement Report 2026-08-26 · Ongoing
- Z-2977-2026openFDA record
Boston Scientific Corporation
Boston Scientific ENROUTE Transcarotid Neuroprotection System, Material Number SR-200-NPS; Percutaneous catheter
Reason: Potential for arterial sheath tip separation or partial tip separation during use.
Device · initiated 2026-07-09 · Enforcement Report 2026-08-26 · Ongoing
- Z-2978-2026openFDA record
Boston Scientific Corporation
Boston Scientific ENROUTE Transcarotid Neuroprotection System Plus, Material Number SR-250-NPS; Percutaneous catheter
Reason: Potential for arterial sheath tip separation or partial tip separation during use.
Device · initiated 2026-07-09 · Enforcement Report 2026-08-26 · Ongoing
- D-0788-2026openFDA record
Fresenius Kabi USA, LLC
Morphine Sulfate Injection, USP, 2 mg /mL, 1 mL single-dose Simplist prefilled syringes, For Intramuscular or Intravenous use, Rx only, Fresenius Kabi, Lake Zurich, IL. Unit of Use NDC Number 76045-004-01; Unit of Sale NDC Number 76045-004-11
Reason: Labeling: Label Mixup: MicroVault labeled as Morphine 2 mg/1 mL, contains a Prefilled Syringe of Dilaudid 0.5 mg/0.5 mL.
Drug · initiated 2026-08-04 · Enforcement Report 2026-08-26 · Ongoing
Class II52 recallsShow all
- Z-2989-2026openFDA record
American Contract Systems Inc
American Contract Systems medical procedure kits and trays labeled as: 1) CRANI PACK - 208400, Model Item Number UDCP57AI; 2) SPINE PACK, Model Item Number HSLM53AC; 3) CRANIAL PACK - 205992, Model Item Number MHCP75AH; 4) IMRI CRANI PACK - 213871, Model Item Number ANCI17AO; 5)…
Reason: The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed¿ Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
Device · initiated 2026-05-19 · Enforcement Report 2026-08-26 · Ongoing
- Z-2990-2026openFDA record
American Contract Systems Inc
American Contract Systems medical procedure kits and trays labeled as: 1) ANTERIOR CERVICAL PACK - 205988, Model Item Number MHAC65AH; 2) SPINE PACK - 206013, Model Item Number MHSP99AF.
Reason: The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed¿ Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
Device · initiated 2026-05-19 · Enforcement Report 2026-08-26 · Ongoing
- D-0791-2026openFDA record
American Regent, Inc.
Nitroglycerin injection, USP, 50 mg/10 mL (5 mg/mL), packaged in a) 10 mL vials (NDC 0517-4810-01), b) 25x10mL vials (NDC 0517-4810-25) RX only, AMERICAN REGENT, INC., Shirley, NY 11967.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as hair, glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- D-0792-2026openFDA record
American Regent, Inc.
Cyanocobalamin injection USP, 1,000 mcg/mL, For IM or SC Use Only, packaged in a) 1 mL Multi-Dose Vial (NDC 0517-0031-01), and b) 25x1 mL Multi-Dose Vials (NDC 0517-0031-25) Rx only, AMERICAN REGENT, INC., SHIRLEY, NY 11967.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as hair, glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- D-0793-2026openFDA record
American Regent, Inc.
Cyanocobalamin injection USP, 10,000 mcg/10 mL (1,000 mcg/mL), For IM or SC Use Only, packaged in a) 10 mL Multi-Dose Vial (NDC 0517-0032-01) and b) 25x10 mL Multi-Dose Vials (NDC 0517-0032-25), Rx Only, AMERICAN REGENT, INC., SHIRLEY, NY 11967.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as hair, glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- D-0794-2026openFDA record
American Regent, Inc.
AMINAOCAPROIC ACID INJECTION, USP, 250 mg/mL (5 g/20 mL), packaged in a) 20 mL MULTIPLE DOSE VIAL FOR IV INFUSION (NDC 0517-9120-01), and b) 25x20 mL MULTIPLE DOSE VIALS (NDC 0517-9120-25), Rx Only, American Regent, Inc., Shirley, NY 11967.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as hair, glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- D-0795-2026openFDA record
American Regent, Inc.
Atropine Sulfate injection, USP, 1 mg/mL, For intravenous Use, Sterile, packaged in a) 1 mL Single-Dose Vial (NDC 0517-1001-01), and b) 25x1 mL Single-Dose Vial (NDC 0517-1001-25), Rx only, AMERICAN REGENT INC., SHIRLEY, NY 11967.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as hair, glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- D-0796-2026openFDA record
American Regent, Inc.
ACETYLCYSTEINE SOLUTION, USP, 10% (100 mg/mL), packaged in a) 4 mL Vial NOT FOR INJECTION (NDC 0517-7504-01), and b) 25x4 mL Vials NOT FOR INJECTION (NDC 0517-7504-25), Rx Only, AMERICAN REGENT, INC., SHIRLEY, NY 11967.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as hair, glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- D-0797-2026openFDA record
American Regent, Inc.
ACETYLCYSTEINE SOLUTION, USP, 20% (200 mg/mL), packaged in a) 4 mL Vial NOT FOR INJECTION (NDC 0517-7604-01), and b) 25x4 mL Vials NOT FOR INJECTION (NDC 0517-7604-25), Rx only, AMERICAN REGENT, INC., SHIRLEY, NY 11967.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as hair, glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- D-0798-2026openFDA record
American Regent, Inc.
Selenious Acid Injection, USP 12 mcg/2 mL (6 mcg/2mL), For intravenous use, packaged in a) 2mL Single-Dose Vial (NDC 0517-6502-01), and b)10x2mL Single-Dose Vial (NDC 0517-6502-10), Rx Only, AMERICAN REGENT, INC., SHIRLEY, NY 11967.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as hair, glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- D-0799-2026openFDA record
American Regent, Inc.
Multrys (Trace Elements Injection 4, USP), For intravenous infusion, packaged in a) 1 mL Single-Dose Vial (NDC 0517-9302-01), and b) 25x1 mL Single-Dose Vials (NDC 0517-9302-25) Rx Only, AMERICAN REGENT, INC., SHIRLEY, NY 11967.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as hair, glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- D-0801-2026openFDA record
American Regent, Inc.
Tralement (trace elements injection 4*, USP), packaged in a) 1mL Single Dose vials (NDC 0517-9305-01) and b) 5x1mL Single Dose vials (NDC 0517-9305-25), Rx only, AMERICAN REGENT, INC., SHIRLEY, NY 11967.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as hair, glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- D-0802-2026openFDA record
American Regent, Inc.
niCARdipine Hydrochloride Injection, USP, 25 mg/10mL (2.5 mg/mL), packaged in a) 10 mL Single Dose Vial (NDC 72572-470-01) and b) 10x10 mL Single Dose Vials (NDC 72572-470-10), Rx Only, Mfd for: Civica, Inc., Lehi, UT, Mfd by: American Regent, Inc., New Albany, OH 43054.
Reason: Presence of Particulate Matter: Product contaminated with particulate matter identified as hair, glass and/or paraformaldehyde
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- Z-2958-2026openFDA record
Arrowhead DE LLC
Arrow-LOK Implant, REF: 21-253519, 21-253522, 21-253525, 21-253528, 21-253530, 21-253533, 21-253535, 21-253540, 21-253545, 21-253550, 21-353513, 21-353516, 27-253519, 27-253522, 27-253525, 27-253528, 27-253530, 27-253533, 27-253535. Implant indicated for fixation of osteotomies,…
Reason: Implant did not undergo required gamma radiation dose audit. A potential reduction in sterility assurance could lead to patient infection.
Device · initiated 2026-06-23 · Enforcement Report 2026-08-26 · Ongoing
- Z-2964-2026openFDA record
AVID Medical, Inc.
Halyard LAVH PACK, Model/Catalog Number: USAR053-11 containing Clearify Visualization System (5303907).
Reason: The kit component was subjected to EO sterilization, contrary to the manufacturer recommendation, which may have led to the component leaking in a kit.
Device · initiated 2026-07-10 · Enforcement Report 2026-08-26 · Ongoing
- Z-2973-2026openFDA record
Beckman Coulter Inc.
Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry A…
Reason: a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently…
Device · initiated 2026-07-15 · Enforcement Report 2026-08-26 · Ongoing
- Z-2974-2026openFDA record
Beckman Coulter Inc.
Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers
Reason: a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently…
Device · initiated 2026-07-15 · Enforcement Report 2026-08-26 · Ongoing
- Z-2961-2026openFDA record
Boston Scientific Corporation
OverStitch Suture Cinch, 1pk. Endoscopic tissue approximation device.
Reason: Device may fail to deploy the cinch implant and/or fail to cut the suture as intended.
Device · initiated 2026-07-14 · Enforcement Report 2026-08-26 · Ongoing
- Z-2962-2026openFDA record
Boston Scientific Corporation
Suture Cinch Long 1pk. Endoscopic tissue approximation device.
Reason: Device may fail to deploy the cinch implant and/or fail to cut the suture as intended.
Device · initiated 2026-07-14 · Enforcement Report 2026-08-26 · Ongoing
- Z-2986-2026openFDA record
ConvaTec, Inc
Esteem Body Soft Convex Urostomy Pouch Model/Catalog Numbers: 423703, 423704, 423705, 423707, 423708, 423709, 423710, 423712, 423713, 423714, 423715, 423716, 423717, 423718, 423721, 423722, 423723, 423725, 423726, 423727.
Reason: Urostomy Pouches may leak from cap during use.
Device · initiated 2026-07-06 · Enforcement Report 2026-08-26 · Ongoing
- Z-2950-2026openFDA record
Covidien, LLC
EndoFLIP System that includes: ENDOFLIP 1 SYSTEM EF-100 05391530810050 10884521809307 REFURBISHED SYSTEM RFG-EF100 ENDOFLIP 1 10884521808096 05391530810166 DL ENDOFLIP 1 10884521786684 Endoscopic measuring system, including catheter components.
Reason: The system's 24-hour usage limit results in catheters incorrectly flagged as expired, making them unavailable for clinical procedures.
Device · initiated 2026-05-13 · Enforcement Report 2026-08-26 · Ongoing
- Z-2983-2026openFDA record
Fort Defiance Industries, LLC
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Reason: The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Device · initiated 2026-07-08 · Enforcement Report 2026-08-26 · Ongoing
- Z-2951-2026openFDA record
Foundation Medicine, Inc.
Hamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LCDx.
Reason: Software issue that can cause an increase in spontaneous software closures.
Device · initiated 2026-06-10 · Enforcement Report 2026-08-26 · Ongoing
- Z-2968-2026openFDA record
Gentherm Medical, LLC
Blanketrol II Model 222S; Thermal regulating system.
Reason: Overheating of the 115V CPU board may cause the device to shut down.
Device · initiated 2026-07-13 · Enforcement Report 2026-08-26 · Ongoing
- Z-2969-2026openFDA record
Gentherm Medical, LLC
Innercool Coolblue Console, Model STx; Thermal regulating system.
Reason: Overheating of the 115V CPU board may cause the device to shut down.
Device · initiated 2026-07-13 · Enforcement Report 2026-08-26 · Ongoing
- Z-2970-2026openFDA record
Gentherm Medical, LLC
Blanketrol III, Model 233; Thermal regulating system.
Reason: Overheating of the 115V CPU board may cause the device to shut down.
Device · initiated 2026-07-13 · Enforcement Report 2026-08-26 · Ongoing
- D-0771-2026openFDA record
Golden State Medical Supply Inc.
Carbamazepine Tablets, USP, 200mg, packaged in a)1000-count bottles (NDC 51407-215-10), b) 100-count bottles (NDC 51407-215-01), Rx Only, Manufactured by: Taro Pharmaceutical Industries Ltd., Marketed by: GSMS, Incorporated, Camarillo, CA 93012 USA
Reason: CGMP deviations: tablets with black specks.
Drug · initiated 2026-08-07 · Enforcement Report 2026-08-26 · Ongoing
- Z-2952-2026openFDA record
Intersurgical Inc
One-Piece Guedel Airway, size 00, ISO 5.0, Blue. Oropharyngeal Airway Device.
Reason: Devices could potentially be occluded or partially occluded.
Device · initiated 2026-06-15 · Enforcement Report 2026-08-26 · Ongoing
- Z-2953-2026openFDA record
Intersurgical Inc
One-Piece Guedel Airway, size 000, ISO 3.5, Light Green. Oropharyngeal Airway Device.
Reason: Devices could potentially be occluded or partially occluded.
Device · initiated 2026-06-15 · Enforcement Report 2026-08-26 · Ongoing
- Z-2957-2026openFDA record
Kico Knee Innovation Company
Brand Name: Kico Knee Innovation Product Name: ARVIS SURGICAL NAVIGATION SYSTEM GEN 2 Model/Catalog Number: IN-29200 Software Version: ARVIS Hip and Knee Gap Balance Software v3.3.2 (IN-28200) ARVIS Shoulder Software v2.0.3 (IN-23630) Product Description: The ARVIS system is a c…
Reason: Due to an internal sensor issue, image displayed with eyepiece may appear tilted relative to the eyepiece housing
Device · initiated 2026-07-13 · Enforcement Report 2026-08-26 · Ongoing
- Z-2963-2026openFDA record
Maquet Critical Care AB
GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilator system, for facility use. Maquet Critical Care AB, is a subsidiary of Getinge.
Reason: The module did not undergo electrical safety tests and functional tests during production.
Device · initiated 2026-06-30 · Enforcement Report 2026-08-26 · Ongoing
- Z-2980-2026openFDA record
Medtronic MiniMed, Inc.
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;
Reason: Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Device · initiated 2026-07-13 · Enforcement Report 2026-08-26 · Ongoing
- Z-2981-2026openFDA record
Medtronic MiniMed, Inc.
MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
Reason: Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Device · initiated 2026-07-13 · Enforcement Report 2026-08-26 · Ongoing
- Z-2982-2026openFDA record
Medtronic MiniMed, Inc.
MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW
Reason: Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Device · initiated 2026-07-13 · Enforcement Report 2026-08-26 · Ongoing
- D-0774-2026openFDA record
Mylan Pharmaceuticals Inc
Acamprosate Calcium, Delayed-Release Tablets, 333 mg, 180 tablets bottles, Rx only, Mylan Pharmaceuticals Inc., Manufactured for: Mylan Pharmaceuticals Inc., Morgantown, WV 266505, NDC 0378-6333-80.
Reason: Failed Dissolution Specifications
Drug · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- Z-2967-2026openFDA record
Paragonix Technologies Inc.
Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001
Reason: Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.
Device · initiated 2026-07-20 · Enforcement Report 2026-08-26 · Ongoing
- Z-2955-2026openFDA record
Philips North America Llc
Cardiac Workstation 5000. Product Number: 860439.
Reason: Device will intermittently provide inaccurate data related to native heart rate.
Device · initiated 2026-07-01 · Enforcement Report 2026-08-26 · Ongoing
- Z-2956-2026openFDA record
Philips North America Llc
Cardiac Workstation 7000. Product Number: 860441.
Reason: Device will intermittently provide inaccurate data related to native heart rate.
Device · initiated 2026-07-01 · Enforcement Report 2026-08-26 · Ongoing
- D-0767-2026openFDA record
Regeneron Pharmaceuticals Inc
Libtayo (cemiplimab-rwlc) Injection, 350 mg/7 mL (50 mg/mL), For Intravenous Infusion After Dilution, Single-Dose Vial, Rx only, Manufactured by: Regeneron Pharmaceuticals, Inc., Tarrytown, NY 10591; Marketed by: Regeneron Pharmaceuticals, Inc. (Tarrytown, NY 10591), NDC 61755-0…
Reason: Lack of Assurance of Sterility
Drug · initiated 2026-07-28 · Enforcement Report 2026-08-26 · Ongoing
- Z-2987-2026openFDA record
restor3d Inc.
Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is…
Reason: Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.
Device · initiated 2026-06-30 · Enforcement Report 2026-08-26 · Ongoing
- Z-2954-2026openFDA record
SDI Limited
Dentsply Sirona Capsule Mixer Model No. 5546068
Reason: Potential for electrical discharge from unit when unplugged.
Device · initiated 2026-07-20 · Enforcement Report 2026-08-26 · Ongoing
- D-0764-2026openFDA record
Shilpa Medicare Limited
PEMRYDI RTU (pemetrexed injection), 100 mg/10 mL (10 mg/mL), Single-dose vial, Rx only, Manufactured by: Zydus Lifesciences Limited, Ahmedabad, India; Distributed by: Amneal Pharmaceuticals LC, Bridgewater, NY 08807. NDC: 70121-2453-1
Reason: Discolored solution. The firm has received market complaints reporting green to dark green discoloration in certain vials
Drug · initiated 2026-07-29 · Enforcement Report 2026-08-26 · Ongoing
- D-0765-2026openFDA record
Shilpa Medicare Limited
PEMRYDI RTU(pemetrexed injection), 500 mg/50 mL (10 mg/mL), Single-dose vial, Rx only, Manufactured by: Zydus Lifesciences Limited, Ahmedabad, India; Distributed by: Amneal Pharmaceuticals LC, Bridgewater, NY 08807. NDC: 70121-2461-1
Reason: Discolored solution. The firm has received market complaints reporting green to dark green discoloration in certain vials
Drug · initiated 2026-07-29 · Enforcement Report 2026-08-26 · Ongoing
- Z-2965-2026openFDA record
Siemens Healthcare Diagnostics Products Ltd.
Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF:L2KSY2(D)/10471981 IVD, RxOnly
Reason: Syphilis Screen assay may contain a negative patient bias resulting in low reactive patient results being reported as nonreactive.
Device · initiated 2026-07-13 · Enforcement Report 2026-08-26 · Ongoing
- Z-2966-2026openFDA record
Siemens Healthcare Diagnostics Products Ltd.
Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF(s):L2kSY2/10489085; L2KSY6/10488319 IVD, RxOnly
Reason: Syphilis Screen assay may contain a negative patient bias resulting in low reactive patient results being reported as nonreactive.
Device · initiated 2026-07-13 · Enforcement Report 2026-08-26 · Ongoing
- Z-2975-2026openFDA record
Siemens Healthcare Diagnostics, Inc.
Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL
Reason: Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadeq…
Device · initiated 2026-07-13 · Enforcement Report 2026-08-26 · Ongoing
- Z-2976-2026openFDA record
Siemens Healthcare Diagnostics, Inc.
Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
Reason: Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadeq…
Device · initiated 2026-07-13 · Enforcement Report 2026-08-26 · Ongoing
- Z-2939-2026openFDA record
Spectral Instruments Inc
Lago X
Reason: Momentary absence of On-Screen "X-Ray On" banner during run readiness check.
Device · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- Z-2940-2026openFDA record
Spectral Instruments Inc
Ami HTX
Reason: Momentary absence of On-Screen "X-Ray On" banner during run readiness check.
Device · initiated 2026-07-24 · Enforcement Report 2026-08-26 · Ongoing
- D-0770-2026openFDA record
SUN PHARMACEUTICAL INDUSTRIES INC
Carbamazepine Tablets, USP, 200mg, 1000 count bottles, Rx only, Manufactured by: Taro Pharmaceutical Industries Ltd., Haifa Bay, Israel 2624761, Distributed by: Taro Pharmaceuticals USA, Inc., Hawthorne, NY 10532, NDC 51672-4005-3.
Reason: CGMP Deviations; black spots found on tablets from burnt excipient during manufacturing
Drug · initiated 2026-08-03 · Enforcement Report 2026-08-26 · Ongoing
- Z-2959-2026openFDA record
Tobii Dynavox Llc
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
Reason: Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Device · initiated 2026-07-08 · Enforcement Report 2026-08-26 · Ongoing
- Z-2960-2026openFDA record
Tobii Dynavox Llc
TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
Reason: Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Device · initiated 2026-07-08 · Enforcement Report 2026-08-26 · Ongoing
Class III2 recallsShow all
- Z-2971-2026openFDA record
Paragon 28, Inc.
Custom Made Device Implant 6123 Navicular Cage Small Right Rev A, Model/Catalog Number 6123S R; foot implant
Reason: Device was recalled due to an elevated oxygen level in excess of the allowable limit. Elevated oxygen levels may contribute to reduced material ductility.
Device · initiated 2026-07-10 · Enforcement Report 2026-08-26 · Ongoing
- Z-2972-2026openFDA record
Paragon 28, Inc.
Custom Made Device Implant 6123 Navicular Cage Nominal Right Rev A, Model/Catalog Number 6123N R; foot implant
Reason: Device was recalled due to an elevated oxygen level in excess of the allowable limit. Elevated oxygen levels may contribute to reduced material ductility.
Device · initiated 2026-07-10 · Enforcement Report 2026-08-26 · Ongoing
New OAI classifications
2 inspections classified Official Action Indicated that FDAgent first recorded from FDA's inspection classification data this week, after its reads of that data in the 8 days before had not found them. Older inspections FDAgent adds to its archive are left out. Dates are inspection end dates. Inspection classifications last synced Nov 23, 2026, 04:25 UTC.
- Inspection ended 2026-05-29FEI 3005334016
Northland Nuclear Medicine, LLCOAI
Bismarck, North Dakota, United States · Drugs · Drug Quality Assurance · recorded 2026-08-25
- Inspection ended 2026-06-09FEI 3009867894
Umicore Argentina S.A.OAI
Pilar, Argentina · Drugs · Drug Quality Assurance · recorded 2026-08-25
Get this page by email every Monday
A free weekly email with each week’s warning letters, new Form 483s, OAI classifications and Class I recalls, counted the same way as this page. Just your email address.
About this data
- Weeks are ISO weeks, Monday to Sunday, in UTC. Each section is dated by its own FDA date: posting date for warning letters, reading-room publish date for Form 483s (each inspection's 483 counts once, when FDA first listed it; FDA's re-posts and amended 483s with the FEI and record date of one it listed before are shown apart), Enforcement Report date for recalls, and the date published FDA prints beside each firm on an import alert. FDA publishes no date for an inspection classification, so an OAI classification counts as new on the day FDAgent first recorded the inspection, and only when FDAgent's complete reads of FDA's inspection data in the 8 days before had covered the inspection's end date without finding it. Older inspections FDAgent adds to its archive, by a backfill or a wider read, never count as new, however late FDA posted them. Inspections FDAgent recorded before it began keeping a record of its reads, in November 2026, count only when they ended in the 13 weeks before, and the Veterinary inspections of FDAgent's first read of them, that same month, never count.
- Import alert entries come from FDA's import alert pages, where FDA lists firms by name and address, not by FEI. FDAgent links an entry to a facility only when FDA printed the firm's FEI, never by name. Earlier weeks list only entries still on FDA's pages, because FDA deletes an entry when it removes a firm, and FDAgent keeps removed entries only from the day it began recording them. FDAgent lists an entry here only when its name shows it is a firm, through a legal form such as Ltd, Inc or GmbH, a business word such as Laboratories or Trading, or a web address, or when FDA's inspection or drug registration records hold a firm under exactly that name; entries that may name a person are shown only to signed-in users.
- Import refusals are not listed by week: FDA reports a refusal weeks after it happens, so a week's count would keep changing after the week closed. Each facility page lists its own refusals.
- Warning letters are those issued by FDA's drug, biologics and device centers and their field and quality offices. Letters from FDA's food, tobacco, veterinary, cosmetics and dietary supplement offices are left out.
- FDA posts only some Form 483s to its Electronic Reading Room, so a missing 483 does not mean none was issued.
- FDA publishes an inspection's classification once it is final and updates its inspection data weekly. Source: FDA inspections FAQ.
- Recalls come from openFDA, which republishes FDA's weekly Enforcement Report; a report's recalls appear here once openFDA publishes it.
- FDAgent is independent of FDA. Every row links to its FDA source; check it before relying on a record.