21 CFR 812.5: Labeling of investigational devices
FDA's inspection data lists 5 Form 483 citations of 21 CFR 812.5 at 5 facilities, from August 2018 to October 4, 2024.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 5
- At 5 inspections
- Facilities cited
- 5
- Distinct FEIs
- Most recent citation
- Oct 4, 2024
- United Health Products, Inc.
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 812.5 from the Electronic Code of Federal Regulations.
§ 812.5 Labeling of investigational devices.
(a) Contents. An investigational device or its immediate package shall bear a label with the following information: the name and place of business of the manufacturer, packer, or distributor (in accordance with § 801.1), the quantity of contents, if appropriate, and the following statement: “CAUTION—Investigational device. Limited by Federal (or United States) law to investigational use.” The label or other labeling shall describe all relevant contraindications, hazards, adverse effects, interfering substances or devices, warnings, and precautions.
(b) Prohibitions. The labeling of an investigational device shall not bear any statement that is false or misleading in any particular and shall not represent that the device is safe or effective for the purposes for which it is being investigated.
(c) Animal research. An investigational device shipped solely for research on or with laboratory animals shall bear on its label the following statement: “CAUTION—Device for investigational use in laboratory animals or other tests that do not involve human subjects.”
(d) The appropriate FDA Center Director, according to the procedures set forth in § 801.128 or § 809.11 of this chapter, may grant an exception or alternative to the provisions in paragraphs (a) and (c) of this section, to the extent that these provisions are not explicitly required by statute, for specified lots, batches, or other units of a device that are or will be included in the Strategic National Stockpile.
21 CFR Part 812: Investigational Device Exemptions, Subpart ARead on eCFR
Citations by year
Form 483 citations of 21 CFR 812.5 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 1 in 2018
| Year | Form 483 citations |
|---|---|
| 2018 | 1 |
| 2019 | 1 |
| 2021 | 1 |
| 2023 | 1 |
| 2024 | 1 |
How FDA words this citation
FDA's short descriptions and citation text on these 5 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Label does not contain required information5 citations
Citation text
- “The label for an investigational device does not include the name and place of business of manufacturer, packer, or distributor, the quantity of contents and the statement "CAUTION -- Investigational device. Limited by Federal (or United States) law to investigational use.".”3 citations
- “The label for an investigational device does not include the name and place of business of manufacturer, packer, or distributor and the statement "CAUTION -- Investigational device. Limited by Federal (or United States) law to investigational use.".”1 citations
- “The label for an investigational device does not include the statement "CAUTION -- Investigational device. Limited by Federal (or United States) law to investigational use.".”1 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 812.5, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
Often cited at the same inspection
Of the 5 inspections where FDA cited 21 CFR 812.5, the share where it also cited each of these sections.
- 21 CFR 812.140Records3 (60%)
- 21 CFR 812.2Applicability3 (60%)
- 21 CFR 812.40General responsibilities of sponsors3 (60%)
- 21 CFR 50.20General requirements for informed consent2 (40%)
- 21 CFR 50.27Documentation of informed consent1 (20%)
- 21 CFR 812.100General responsibilities of investigators1 (20%)
Ask FDAgent about 21 CFR 812.5
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 812.5; FDAgent quotes that sentence exactly.