21 CFR 807.81: When a premarket notification submission is required
FDA's inspection data lists no Form 483 citations of 21 CFR 807.81. 3 warning letters state a violation of it.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 0
- At 0 inspections
- Facilities cited
- 0
- Distinct FEIs
- Most recent citation
- —
- None
- Warning letters
- 3
- State a violation of it
The regulation
The current text of 21 CFR 807.81 from the Electronic Code of Federal Regulations.
§ 807.81 When a premarket notification submission is required.
(a) Except as provided in paragraph (b) of this section, each person who is required to register his establishment pursuant to § 807.20 must submit a premarket notification submission to the Food and Drug Administration at least 90 days before he proposes to begin the introduction or delivery for introduction into interstate commerce for commercial distribution of a device intended for human use which meets any of the following criteria:
(1) The device is being introduced into commercial distribution for the first time; that is, the device is not of the same type as, or is not substantially equivalent to, (i) a device in commercial distribution before May 28, 1976, or (ii) a device introduced for commercial distribution after May 28, 1976, that has subsequently been reclassified into class I or II.
(2) The device is being introduced into commercial distribution for the first time by a person required to register, whether or not the device meets the criteria in paragraph (a)(1) of this section.
(3) The device is one that the person currently has in commercial distribution or is reintroducing into commercial distribution, but that is about to be significantly changed or modified in design, components, method of manufacture, or intended use. The following constitute significant changes or modifications that require a premarket notification:
(i) A change or modification in the device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process.
(ii) A major change or modification in the intended use of the device.
(b)(1) A premarket notification under this subpart is not required for a device for which:
(i) A premarket approval application under section 515 of the act, or for which a petition to reclassify under section 513(f)(2) of the act, is pending before the Food and Drug Administration, or
(ii) There is a predetermined change control plan (PCCP) cleared under section 515C of the act, provided that the change is consistent with the PCCP.
(2) The appropriate FDA Center Director may determine that the submission and grant of a written request for an exception or alternative under § 801.128 or § 809.11 of this chapter satisfies the requirement in paragraph (a)(3) of this section.
(c) In addition to complying with the requirements of this part, owners or operators of device establishments that manufacture radiation-emitting electronic products, as defined in § 1000.3 of this chapter, shall comply with the reporting requirements of part 1002 of this chapter.
21 CFR Part 807: Establishment Registration and Device Listing for Manufacturers and Initial Importers of Devices, Subpart ERead on eCFR
Warning letters stating a violation of 21 CFR 807.81
3 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“You currently have clearance, dated August 29, 2011, for a version of the Medfusion® Model 4000 Syringe Infusion Pump which was cleared under K111386 as a Class II device under 21 CFR 880.5725. You also currently have clearance, dated February 1, 2013, for a version of the CADD® Solis VIP Ambulatory Infusion Pump which was cleared under K111275 as a Class II device under 21 CFR 880.5725. However, your firm has failed to submit a premarket notification submission to FDA for significant changes or modifications that could significantly affect the safety or effectiveness of these devices, as required by 21 CFR 807.81(a)(3)(i).”
- ICU MedicalIssued 2025-04-04
“Your Salivary Cortisol ELISA RUO appears to be intended for the quantitative in vitro diagnostic measurement of active free cortisol (hydrocortisone and hydroxycorticosterone) in saliva. You were required to submit to FDA a 510(k) for this device, because either it is a new device (21 CFR 807.81(a)(1))”
- DRG Instruments GmbHIssued 2025-03-31
“Your iPLEX HS Colon Panel intended for use with the MA4 appears to be a tumor profiling test intended for the diagnosis of colorectal cancer. You were required to submit to FDA a 510(k) for this device, because it either is a new device (21 CFR 807.81(a)(1))”
- Agena Bioscience, Inc.Issued 2024-03-21
Ask FDAgent about 21 CFR 807.81
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 807.81; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.